User Manual

3 Monitoring
Monitoring
LIFEPAK 20e Defibrillator/Monitor Operating Instructions 3-17
©2006-2013 Physio-Control, Inc.
How Capnography Works
An EtCO2 sensor continuously monitors carbon dioxide (CO2) that is inspired and exhaled by
the patient. The sensor employs Microstream non-dispersive infrared (IR) spectroscopy to
measure the concentration of CO2 molecules that absorb infrared light.
The CO2 FilterLine system delivers a sample of the exhaled gases directly from the patient into
the CodeManagement Module for CO2 measurement. The low sampling flow rate (50 ml/min)
reduces liquid and secretion accumulation and prevents obstruction, which maintains the shape
of the CO2 waveform.
The CO2 sensor captures a micro sample (15 microliters). This extremely small volume allows
for fast rise time and accurate CO2 readings, even at high respiration rates.
The Microbeam IR source illuminates the sample cell and the reference cell. This proprietary IR
light source generates only the specific wavelengths characteristic of the CO2 absorption
spectrum. Therefore, no compensations are required when concentrations of O2, anesthetic
agent, or water vapor are present in the exhaled breath.
The LIFEPAK 20e defibrillator/monitor is set at the factory to use the capnography Body
Temperature Pressure Saturated (BTPS) conversion method. This option corrects for the
difference in temperature and moisture between the sampling site and alveoli. The correction
formula is 0.97 x the measured EtCO2 value. The BTPS conversion can be turned off in the
Setup Options. Refer to "CO2 Setup Menu," page 8-10 for more information.
EtCO2 Monitoring Waveform Analysis
Valuable information concerning the patient’s expired CO2 can be acquired by examination and
interpretation of the waveform.
The Phases of the Waveform
Figure 3-5 is a graphic representation of a normal capnograph waveform. Four phases of the
waveform require analysis. The flat I–II baseline segment (Respiratory Baseline) represents
continued inhalation of CO2-free gas. This value normally is zero. The II–III segment (Expiratory
Upstroke), a sharp rise, represents exhalation of a mixture of dead space gases and alveolar
gases from acini with the shortest transit times. Phase III–IV (Expiratory Plateau) represents the
alveolar plateau, characterized by exhalation of mostly alveolar gas. Point IV is the end-tidal
(EtCO2) value that is recorded and displayed by the monitor. Phase IV–V (Inspiratory
Downstroke), a sharp fall, reflects the inhalation of gases that are CO2-free. Alterations of the
normal capnograph or EtCO2 values are the result of changes in metabolism, circulation,
ventilation, or equipment function.
Figure 3-5 Phases of the Respiratory Waveform
Infection hazard
Do not reuse, sterilize, or clean Microstream
®
CO2 accessories as they are designed for single-
patient one-time use.
WARNINGS! (CONTINUED)
IV
V
II
III
I