4 Therapy
Therapy
LIFEPAK 20e Defibrillator/Monitor Operating Instructions 4-7
©2006-2013 Physio-Control, Inc.
6 Press the ANALYZE button to initiate analysis. Stop CPR.
7 Follow screen messages and voice prompts provided by the AED.
The SAS analyzes the patient's ECG and advises either
SHOCK ADVISED or NO SHOCK
ADVISED
.
Shock Advised
The PUSH ANALYZE message and voice
prompt occur when the patient is
properly connected to the AED. The
PUSH ANALYZE message will stay on the
screen and the analyze LED flashes until
ANALYZE is pressed.
You will see and hear ANALYZING NOW,
STAND CLEAR
. Do not touch or move
the patient or therapy cable during
analysis. ECG analysis requires
approximately 6 to 9 seconds. The
analyze LED illuminates during analysis.
If the AED detects a shockable ECG
rhythm, you will see and hear
SHOCK
ADVISED
. The AED begins charging to
the setup joule setting for shock #1. A
rising tone indicates that the AED is
charging.
When charging is complete, the AED
displays the available energy.
You will see and hear STAND CLEAR,
PUSH TO SHOCK
( ) followed by a
“shock ready” tone. The shock LED
flashes.
Clear everyone away from the patient,
bed, or any equipment connected to the
patient.
Press the button to discharge the
AED.