4-10 LIFEPAK 20e Defibrillator/Monitor Operating Instructions
Motion Detected
Continuous Patient Surveillance System
Special AED Setup Options
The following descriptions of voice prompts and messages explain special cprMAX technology
setup options (refer to Appendix F).
Initial CPR - CPR First
When the INITIAL CPR option is set to CPR FIRST, you will be prompted to START CPR
immediately after the AED is turned on.
If motion is detected during the ECG
analysis, you will see and hear
MOTION
DETECTED, STOP MOTION
followed by a
warning tone. Analysis is inhibited up to
10 seconds. After 10 seconds, even if
motion is still present, the analysis
proceeds to completion. Refer to Table
4-1 for possible motion causes and
suggested solutions.
When the device is not analyzing the
ECG and is not in CPR Time, the
Continuous Patient Surveillance System
(CPSS) actively monitors the ECG for a
potentially shockable rhythm.
If a shockable rhythm is detected, you
will see and hear
PUSH ANALYZE.
You should:
1 Confirm the patient is unconscious,
pulseless, and not breathing
normally.
2 Confirm no motion is present. Stop
CPR.
3 Press
ANALYZE.
The AED begins to analyze the patient’s
ECG.
For information about changing the
defibrillation mode, refer to Section 8.
You will see and hear START CPR.