LIFEPAK 20e Defibrillator/Monitor Operating Instructions 4-13
©2006-2013 Physio-Control, Inc.
Troubleshooting Tips for AED Mode
Table 4-1 Troubleshooting Tips for AED Mode
Observation Possible Cause Corrective Action
1 CONNECT ELECTRODES
message appears.
Inadequate connection to
defibrillator.
Check for electrode
connection.
Electrodes do not adhere
properly to the patient.
Press electrodes firmly on
patient’s skin.
Clean, shave, and dry the
patient’s skin as
recommended.
Electrodes are dry, damaged,
or out of date.
Replace the electrodes.
2
REMOVE TEST PLUG
message appears.
Test plug connected to therapy
cable.
Disconnect test plug from
therapy cable.
Connect electrodes to
therapy cable.
3
MOTION DETECTED and
STOP MOTION messages
appear during analysis.
Patient movement. Stop CPR during analysis.
When patient is being
manually ventilated, press
ANALYZE after complete
exhalation
.
Move patient to stable
location when possible.
Patient movement because of
agonal respirations.
Press
ANALYZE immediately
after exhalation or wait until
agonal respirations are
slower or absent.
Electrical/radio frequency
interference.
Move hand-held
communication devices or
other suspected devices
away from the defibrillator
when possible.
4
DISARMING message
appears.
Therapy cable or electrodes
disconnected from AED or
patient.
Confirm secure connection
and press
ANALYZE.
Shock button not pressed
within 60 seconds or door is
open.
Press
ANALYZE again.
Press
SHOCK immediately
when directed.
5 Voice prompts sound faint
or distorted.
Low battery power. Connect to AC power.
6
LOW IMPEDANCE–
RECHARGING
message
appears.
Patient impedance <15 ohms
detected.
No action required.
7 Metronome not heard. Metronome only sounds during
CPR time.
No action required.
Metronome not set to be
ON in
AED mode.
Refer to "AED Mode Setup
Menu," page 8-7.