5 Product information
Standard oscilloscope connectors
The PicoScope 3000 Series PC Oscilloscopes have input channels with standard BNC
connectors and standard input impedance. They are therefore compatible with most
oscilloscope probes including x10 and switched x1/x10 types.
The probes supplied with the PicoScope 3000 A and B Series oscilloscopes and MSOs
have been trimmed specifically for use with the scopes they are supplied with. For
optimum performance, please use the probes supplied. Although other oscilloscope
probes can be used, the specified performance cannot be guaranteed. Replacement
probes matched to your PicoScope 3000 can be ordered from Pico Technology.
Signal generator (GEN) output
The GEN connector carries the output of the oscilloscope's built-in signal generator,
which can generate a number of built-in waveforms, as well as user-defined arbitrary
waveforms (B/MSO Models only).
If you are using the PicoScope 6 program, refer to the PicoScope 6 User's Guide for
information on how to configure the signal generator.
If you are writing your own software, refer to the PicoScope 3000 Series (A API)
Programmer's Guide.
External trigger (EXT) input (not MSOs)
The EXT input can be used as a trigger source. It is selected using the trigger drop-
down menu in the PicoScope software, or using a function call if you are writing your
own software.
The EXT input uses dedicated circuitry with a software-configurable threshold to
detect a trigger signal. This has the advantage of freeing the analog channels for
viewing signals. The input characteristics of the EXT input have been matched to a
scope channel so that a supplied probe can be trimmed on one of the channels and
then used with the EXT input to give best vertical accuracy. If trigger timing accuracy
is critical, we recommend using one of the main input channels as the trigger source.
These channels use accurate digital triggering (to one sample period) and have a
vertical resolution of 1 LSB.