This product is certified to conform to the Class IIa requirements of the 93/42/EEC Medical Devices Directive. Install & Service Manual Prudent Flat-panel Digital X-ray Detector Model: Prudent1717/1417/1212 ver. 1.0.8 Be sure to read and understand this manual thoroughly before using the product, and keep this manual in an easily accessible location for quick reference when required. ⓒ 2020. PIXXGEN Corporation. All rights reserved.
Install & Service Manual AT-SM-04 Prudent │ Flat-panel Digital X-ray Detector Rev.1.0.8 INTRODUCTION PRUDENT is a high-resolution digital X-ray imaging device commonly referred to as a flat panel detect. A built-in a-Si TFT flat panel type digital sensor receives X-ray and converts to digital image. X-ray photons are converted to digital output signals. The digital signals are then read out by TFTs.
Install & Service Manual AT-SM-04 Prudent │ Flat-panel Digital X-ray Detector Rev.1.0.8 REVISION HISTORY PRUDENT Manual Revision History Revision Revisions Revised Date Reviser MoonKi Choi 1.0.0 Initial drafting 2015.03.20 NamYeol Lee JiHong Jung 1.0.1 Add related content: Introduction, Warning, Micromagnetic ect. 2015.04.09 NamYeol Lee 1.0.2 Modified layout: 6.2 Environment, 6.4 PICB 2015.04.14 NamYeol Lee 1.0.3 Add related content: FCC and SAR test 2015.05.06 JiHong Jung 1.0.
Install & Service Manual AT-SM-04 Prudent │ Flat-panel Digital X-ray Detector Rev.1.0.8 Contents INTRODUCTION ....................................................................................................................................................... 2 ATTENTION ................................................................................................................................................................ 2 REVISION HISTORY ......................................................
Install & Service Manual AT-SM-04 Prudent │ Flat-panel Digital X-ray Detector Rev.1.0.8 1. Safety Information 1.1. Safety Information PRUDENT passed all legal safety requirements as an electromedical equipment to guarantee the user’s safety. Do not ignore following cautions while handling the products, and read thoroughly this page before use. CAUTION!. ◆ Do not use the instrument, if a malfunction has occurred, until the problems are solved by qualified personnel.
Install & Service Manual AT-SM-04 Prudent │ Flat-panel Digital X-ray Detector Rev.1.0.8 CAUTION!. FCC Statement Federal Communication Commission Interference Statement This equipment has been tested and found to comply with the limits for a Class A digital device, pursuant to part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference when the equipment is operated in a commercial environment.
Install & Service Manual AT-SM-04 Prudent │ Flat-panel Digital X-ray Detector Rev.1.0.8 CAUTION!. IMPORTANT NOTE: FCC Radiation Exposure Statement: This device meets FCC limits for exposure to radio waves. It is designed not to exceed the limits for exposure to radio waves (radio frequency electromagnetic field) adopted by the Federal Communications Commission. This equipment should be installed and operated with minimum distance 0mm between the radiator & your body.
Install & Service Manual AT-SM-04 Prudent │ Flat-panel Digital X-ray Detector Rev.1.0.8 WARNING!. ◆ To avoid risk of electric shock, this equipment must only be connected to a supply mains with protective earth ◆ Do not touch signal input, signal output or other connectors, and the patient simultaneously. ◆ External equipment intended for connection to signal input, signal output or other connectors, shall comply with relevant IEC Standard (e.g.
Install & Service Manual AT-SM-04 Prudent │ Flat-panel Digital X-ray Detector Rev.1.0.8 1.2 Marks Attention, see instruction manual. The CE mark shows that the instrument obtained EU countries’ requirements. CE number can be followed by the CE mark. Manufacturer’s name and address. EU representative’s name, address and contact details. Consult instructions for use Fragile, handle with care This symbol shall be accompanied by the manufacturer's serial number. Temperature limit.
Install & Service Manual AT-SM-04 Prudent │ Flat-panel Digital X-ray Detector Rev.1.0.8 1.3 Condition for installation • Do not install this equipment in any of the locations listed below. - Where the power supply is unstable. - Where temperature and humidity are high. - Where the room is without air-conditioner or ventilation. - Where it is exposed to direct sunlight. • It is highly recommended to separate X-ray power in the distributing board of the electricity of the building.
Install & Service Manual AT-SM-04 Prudent │ Flat-panel Digital X-ray Detector Rev.1.0.8 2. Detector Installation 2.1. Wired Detector 2.1.1. Network Setting The settings between Prudent and Workstation (1) Open the network and sharing center. (2) Open the Network Adaptor properties ⓒ 2020. PIXXGEN Corporation. All rights reserved.
Install & Service Manual AT-SM-04 Prudent │ Flat-panel Digital X-ray Detector Rev.1.0.8 (3) Open TCP/IPv4 and set the static IP as 192.168.100.101~150. 192.168.100.111 is the wired network IP factory default of single detector. 2.1.2. Initial Connection Setting Check Show Network Window message before connecting. Click Connect and then IP and PORT window will appear. Select network window This is to set up connecting ports between PIXX and the workstation computer.
Install & Service Manual AT-SM-04 Prudent │ Flat-panel Digital X-ray Detector Rev.1.0.8 2.1.3. Multi Connection Setting A setting method for using more than two detectors. 192.168.100.111 is the wired network IP. (Factory default of single detector) Single Dual Triple Detector IP 192.168.100.111 192.168.101.112 192.168.102.113 Port 7000 7000 7000 Network Adaptor IP 192.168.100.101 192.168.101.102 192.168.102.
Install & Service Manual AT-SM-04 Prudent │ Flat-panel Digital X-ray Detector Rev.1.0.8 2.2. Wireless Detector 2.2.1. The settings between Bridge (Repeater) and PC (1) Open a web browser (e.g., Internet Explorer, Chrome, Firefox, or Safari) and enter http://tplinklogin.net/ and then login to the system as an administrator (ID: admin / PW: admin). The recommended product for Bridge of Wi-Fi Router is TP Link Archer C7. In CAUTION! this instruction, it has been instructed based on Archer C7 product.
Install & Service Manual AT-SM-04 Prudent │ Flat-panel Digital X-ray Detector Rev.1.0.8 (2) Go to the Advanced > Network > LAN, and change the IP Address to 192.168.120.2 in order to prevent IP conflicts and then click the “Save” button. (3) Go to the Advanced> Wireless Settings, and change the “Wireless Network Name” to “PIXXAP”. And select “Enable WDS Bridging” And then click the “Survey” button. ⓒ 2020. PIXXGEN Corporation. All rights reserved.
Install & Service Manual AT-SM-04 Prudent │ Flat-panel Digital X-ray Detector Rev.1.0.8 (4) Select “PIXXAP120” SSID from AP list, click the “Choose” button. (5) Input the password “1234567890”and then click the “Save” button and reboot. ⓒ 2020. PIXXGEN Corporation. All rights reserved.
Install & Service Manual AT-SM-04 Prudent │ Flat-panel Digital X-ray Detector Rev.1.0.8 (6) Open the network and sharing center. (7) Open the Network Adaptor properties ⓒ 2020. PIXXGEN Corporation. All rights reserved.
Install & Service Manual AT-SM-04 Prudent │ Flat-panel Digital X-ray Detector Rev.1.0.8 (8) Open TCP/IPv4 and set the static IP as 192.168.120.3~254. 2.2.2. Network adapter setting-Network Adaptor (Windows 7,8,10) (1) Installation - Insert driver CD - Run install file following each OS version ⓒ 2020. PIXXGEN Corporation. All rights reserved.
Install & Service Manual AT-SM-04 Prudent │ Flat-panel Digital X-ray Detector Rev.1.0.8 2.2.3. Advanced Network Set-up This is to secure data streaming from disconnection. For the user using own network adapter, it is essential to update the driver of network adapter and adjust the advanced network setting. CAUTION! Realtek network adapter in particular is sensitive to set-up. This set-up is to protect the network from freezing or slowing down.
Install & Service Manual AT-SM-04 Prudent │ Flat-panel Digital X-ray Detector Rev.1.0.8 (2) Run Unidentified Network and proceed to Properties > Composition. (3) Proceed to Driver tab and check the version of current network adapter. (4) The old version of network adapter driver that is already installed on the computer may influence the speed of network system. ⓒ 2020. PIXXGEN Corporation. All rights reserved.
Install & Service Manual AT-SM-04 Prudent │ Flat-panel Digital X-ray Detector Rev.1.0.8 (5) Update the network adapter driver. The information of network adapter driver is available on the manufacturers’ web sites below. http://www.intel.com Intel Tip!. http://www.realtek.com.tw/ Realtek http://www.broadcom.com/ Broad.com. (6) Uncheck all lists in the Power Management option. (7) Disable ‘Wake on~’ and ‘WOL’ options in the Advanced option. ⓒ 2020. PIXXGEN Corporation. All rights reserved.
Install & Service Manual AT-SM-04 Prudent │ Flat-panel Digital X-ray Detector Rev.1.0.8 (8) Change Receive Buffers value to Maximum value. (Receive Buffers setting option of network adapter can be found in Properties of Advanced/ Performance Option.) ⓒ 2020. PIXXGEN Corporation. All rights reserved.
Install & Service Manual AT-SM-04 Prudent │ Flat-panel Digital X-ray Detector Rev.1.0.8 Appendix – Guidelines for Pediatric Subjects 1) Positioning the pediatric patient Pediatric patients are not as likely as adults to understand the need to remain still during the procedure. Therefore it makes sense to provide aids to maintaining stable positioning. It is strongly recommended the use of immobilizing devices such as bean bags and restraint systems(foam wedges, adhesive tapes, etc.
Install & Service Manual AT-SM-04 Prudent │ Flat-panel Digital X-ray Detector Rev.1.0.8 4) Dosimetry Table 2 summarizes the key dosimetry parameters for the four types of radiographic examination for patients ranging from newborn to the adult. In each cell, the first value is the entrance skin air kerma(free-in-air) in u㏉. The second term gives the energy imparted to the patient, expressed in uJ. In parentheses on the second line are the corresponding values of patient effective dose in uSv.
Install & Service Manual AT-SM-04 Prudent │ Flat-panel Digital X-ray Detector Rev.1.0.8 Levels of Ionizing Radiation.2006. Health risks from exposure to low levels of ionizing radiation: BEIR VII phase 2. Washington, D.C.: National Academy of Sciences, National Academies Press.] 4. The following reference gives current data for anteroposterior and transverse body diameter for pediatric patients ranging in age from 0.5 to 20 years: Kleinman, P.L.,K.J. Strauss, D. Zurakowski, K.S. Buckley, and G.A.
Install & Service Manual AT-SM-04 Prudent │ Flat-panel Digital X-ray Detector Rev.1.0.8 Appendix - Check Vaccine programs in your Workstation PC. Some vaccine programs might block detector signal since they recognize our software as a virus or malware. We strongly recommend removing vaccine programs except MS Essential. When using MS Essential, please set up “Excluded files and locations” as shown below. ⓒ 2020. PIXXGEN Corporation. All rights reserved.
Install & Service Manual AT-SM-04 Prudent │ Flat-panel Digital X-ray Detector Rev.1.0.8 Appendix - Virtual Memory / DEP Virtual Memory Windows 7 ⓒ 2020. PIXXGEN Corporation. All rights reserved.
Install & Service Manual AT-SM-04 Prudent │ Flat-panel Digital X-ray Detector Rev.1.0.8 ⓒ 2020. PIXXGEN Corporation. All rights reserved.
Install & Service Manual AT-SM-04 Prudent │ Flat-panel Digital X-ray Detector Rev.1.0.8 DEP Setup in Windows 7 ⓒ 2020. PIXXGEN Corporation. All rights reserved.
Install & Service Manual AT-SM-04 Prudent │ Flat-panel Digital X-ray Detector Rev.1.0.8 ⓒ 2020. PIXXGEN Corporation. All rights reserved.
Install & Service Manual AT-SM-04 Prudent │ Flat-panel Digital X-ray Detector Rev.1.0.8 DEP Setup in Windows 8 - This function may not work exactly. - There are CMOS disturbs this function to use, at that case you need to set DEP release in OS. (1). System Control Panel ➔ System and Security ➔ System (2). Click the Advanced system Setting and then the System Properties window will be shown. And then move the Advanced Tab and click the Settings button of Performance ⓒ 2020. PIXXGEN Corporation.
Install & Service Manual AT-SM-04 Prudent │ Flat-panel Digital X-ray Detector Rev.1.0.8 (3). When the Performance Options window show, move the Data Execution Prevention Tab and then select ‘Turn on DEP for all programs and services except those I select’ (4). Click the ‘Add…’ button and select two files as like below ZView.exe WBMaker.exe ⓒ 2020. PIXXGEN Corporation. All rights reserved.
Install & Service Manual AT-SM-04 Prudent │ Flat-panel Digital X-ray Detector Rev.1.0.8 (5). Confirm the ‘ZView and ‘WBMaker’ was included in List and then click ‘Apply’ button ⓒ 2020. PIXXGEN Corporation. All rights reserved.
Install & Service Manual AT-SM-04 Prudent │ Flat-panel Digital X-ray Detector Rev.1.0.8 Appendix - How to Disable Driver Signature Verification Press the Windows key + C keyboard combination to bring up the Charms Bar, and then click on the Settings Charm. We need to head into the Modern Control Panel, so go ahead and click on the Change PC settings link. When the Control Panel opens, switch over to the “Update & recovery” section. ⓒ 2020. PIXXGEN Corporation. All rights reserved.
Install & Service Manual AT-SM-04 Prudent │ Flat-panel Digital X-ray Detector Rev.1.0.8 Then click on the Recovery option on the left hand side. Once selected, you will see an advanced startup section appear on the right hand side. You will need to click on the “Restart now” button. Once your Computer has rebooted you will need to choose the Troubleshoot option. Then head into Advanced options. ⓒ 2020. PIXXGEN Corporation. All rights reserved.
Install & Service Manual AT-SM-04 Prudent │ Flat-panel Digital X-ray Detector Rev.1.0.8 Then Startup Settings. Since we are modifying boot time configuration settings, you will need to restart your Computer one last time. Finally, you will be given a list of startup settings that you can change. The one we are looking for is “Disable driver signature enforcement”. To choose the setting, you will need to press the F7 key. That’s all there is to it.
Install & Service Manual AT-SM-04 Prudent │ Flat-panel Digital X-ray Detector Rev.1.0.8 PIXXGEN Corporation 5F, SMARTBAY, 123, Beolmal-ro, Dongan-gu Anyang-si Gyeonggi-do, 431-804 Korea Tel: +82-70-4846-8888 FAX: +82-2-6455-2905 Web: http://www.pixxgen.com CMC Medical Devices & Drugs S.L. C/Horacio Lengo Nº 18, CP 29006, Málaga, Spain. ⓒ 2020. PIXXGEN Corporation. All rights reserved.