ASSIGN DIGITAL TRIGGER 
In the ASSIGN DIGITAL TRIGGER options it is possible to select if the add-on should find 
all available muscle activations or if it should find activations depending on digital input 
and output signals, i.e. trigger signals. 
The FIND ACTIVATIONS FROM DIGITAL ON TO DIGITAL OFF option will find all activations 
within the time frames where digital inputs have been received, for example, during the 
time frame a pushbutton has been being pressed. 
The FIND ACTIVATIONS FROM DIGITAL ON TO … SEC will find all activations within the 
specified duration after the first digital inputs received, for example, beginning at the 
moment a pushbutton has been pressed. 
ADVANCED OPTIONS - Filters 
The advanced options allow you to change the parameters which are being used in the 
activation detection algorithm to adjust the processing to your acquisition. In this option, 
you can select if you want to use a lowpass, highpass or a highpass to be applied on your 
signal and set the respective cutoff frequencies for the selected filter. 
To  use  a  lowpass  filter,  click  on  the  LOWPASS  button  and  enter  the  preferred  cutoff 
frequency for this filter in the CUTOFF – LOW field and enter the order of the filter in the 
ORDER field. 
To  use  a  highpass filter,  click on  the HIGHPASS  button and  enter the  preferred cutoff 
frequency for this filter in the CUTOFF – HIGH field and enter the order of the filter in the 
ORDER field. 
To use a bandpass filter, click on the BANDPASS button and enter the preferred cutoff 
frequency for the lowpass component of this filter in the CUTOFF – LOW field and enter 
the preferred cutoff frequency for the highpass component of this filter in the CUTOFF – 
HIGH. Enter the order of the filter in the ORDER field. 
ADVANCED OPTIONS – Process X Channel 
This option allows you to apply the statistical algorithms of this add-on to a selected 
channel  X  within  the  same  time  intervals  of  another  channel  without  any  activation 
dependency  of  channel  X.  The  add-on  will  search  for  activations  on  a  selected  EMG 
channel (click on the channel to select) and analyze the selected channel X at the same 
time intervals where activations have been found in the selected EMG channel even if 
there are no activations or activation triggering activities within these time intervals in 
channel X (e.g. muscle is relaxed in during this interval). 
With this option it is possible to compare two EMG signals within the same time intervals 
without any activation dependencies. 