Specifications PHV1000
Electrical Characteristics
Input Resistance (System) 50 MΩ ± 1 %
Input Capacitance (System) 7.5 pF (typical)
Compensation Range 10 pF - 50 pF (typical)
Input coupling of the measuring instrument 1 MΩ AC / DC
Input Impedance
Note that the Input Impedance of
the probe decreases as the frequency
of the applied signal increases.
Mechanical Characteristics
Weight (probe only ) 67 g
Cable Length 2 m
Probe Tip Diameter 5 mm
Environmental Specifications
Altitude operating up to 2000 m
non-operating up to 15000 m
Temperature Range operating 0° C to +50° C
non-operating -40° C to +71° C
Maximum Relative Humidity operating 80 % relative humidity for temperatures up to +31° C,
decreasing linearly to 40 % at +50° C