User Manual

8
Specifications PHV1000
Specifications that are not defined to be guaranteed are typical and are published as general information to the
user. The instrument should have warmed-up for at least 20 minutes and the environmental conditions do not
exceed the probe `s specified limits.
Electrical Specifications
Attenuation Ratio
(1)
100:1 ± 2 % at DC
Voltage Coefficient 0.0005 %/V (typical)
System Bandwidth 400 MHz (-3 dB)
Probe Risetime 900 ps (10 % - 90 %) (typical)
Maximum Rated Input Voltage
(2)
Measurement category I: 1000 V rms
4000 V
transient overvoltage
Measurement category II: 1000 V rms CAT II
Pollution Degree 2
Voltage Derating
Note that the max. input voltage
rating of the probe decreases as
the frequency of the applied
signal increases.
(1) Connect to oscilloscope with a input impedance of 1 M ± 1 %.
(2) As defined in IEC 61010-031. See definitions explained on page 4.
Typical Voltage Derating PHV1000
Measurement Category I
0
200
400
600
800
1000
1200
0,1 1 10 100 1000
Frequency [MHz]
Amplitude AC rms [V]Sinus