Datasheet

9
Small size
The compact design with only 2.5 mm housing diameter
at the probe tip is ideal for measurements of SMT com-
ponents. It provides a much better visibility over the de-
vice under test than conventional 5 mm housing designs.
A true PMK feature the exchangeable tip is also available
for the PML series. The gold-plated spring tip or the solid
tip are only 0.5 mm in diameter. The tip replacement is
easy and gives the engineer the best choice.
Low Input Capacitance
The standard 10:1 model with an input capacitance of less than 9.5 pF
is most appropriate for measurements of HF signals where a too high
load of the source signal can falsify the result and with particularly sen-
sitive SMT devices it can result in an „Overload“. A system input capaci-
tance (oscilloscope and probe) of less than 5.6 pF is available with the
20:1 model.
True Signal Acquisition on ICs
Particularly for HF measurements on ICs a short as possible ground con-
nection is recommendable. Due to long ground leads most conventional
adaptions bring additional inductance and resonances into the measurement
which will result in false or inaccurate readings. The innovative IC contacting
system of the PML series with 5 different IC adapters from 0.5 to 1.27 mm
pitch in combination with the Ground Blade and IC ground copper pad is ideal
for shortcircuit-safe, reproducible and authentic measurements.
PML High Z Probes
Model PML 211-RO PML 311-RO PML 701-RO PML 711A-RO (***) PML 721-RO
Order-No. 855-211-001 855-311-001 855-701-00E 855-711-A01 855-721-004
Attenuation 10:1 10:1 1:1 10:1 20:1
Input Impedance 10 MΩ 10 MΩ 10 MΩ 20 MΩ
10 pF 10 pF 39 pF (+ Oszi.) 9.5 pF 5.6 pF
Input Coupling of the Measuring Instrument 1 MΩ AC/DC 1 MΩ AC/DC 1 MΩ AC/DC 1 MΩ AC/DC 1 MΩ AC/DC
System Bandwidth (-3 dB) 200 MHz 350 MHz 38 MHz (****) 500 MHz 500 MHz
Risetime (10% - 90 %) 1.4 nsec 1 nsec 9 nsec 700 psec 700 psec
Compensation Range (*) 10 – 25 pF 10 – 25 pF 10 – 25 pF 7 – 20 pF
Rated Voltage 300 V CAT II 300 V CAT II 55 V CAT II 300 V CAT II 300 V CAT II
Pollution Degree 2 2 2 2 2
Safety Standard (IEC / EN) 61010-031 61010-031 61010-031 61010-031 61010-31
Cable Length 1.3 m 1.3 m 1.3 m (**) 1.3 m (**) 1.4 m (**)
(*) Other compensation ranges on request
(**) Further standard cable length 2 m, other cable lengths on request
(***) For current oscilloscope models of different manufacturers we offer performance optimized models.
Please refer to our „probe to scope“ reference chart on our website