Electrical specifications
Attenuation ratio (switchable)
(1)
1:1 10:1 ± 2 % at DC
(at 10:1)
Voltage coefficient – 0.0025 % / V (typical)
System bandwidth 20 MHz
(2)
350 MHz (-3 dB)
Probe risetime 17.5 nsec 1 nsec (10 % – 90 %)
(typical)
Maximum rated input voltage
(3)
Measurement category I: – 400 V rms
1250 V transient overvoltage
Measurement category II: 55 V rms CAT II 300 V rms CAT II
Pollution degree 2
Voltage derating
(1) Connect to oscilloscope with an input impedance of 1 MΩ ± 1 %.
(2) at 15 pF input capacitance of the measuring instrument.
(3) As defined in IEC 61010-031.
Note that the max. input voltage
rating of the probe decreases as
the frequency of the applied signal
increases.
Typical voltage derating PMTG321A (switched to 10:1)
Measurement category I
0
100
200
300
400
0,1 1 10 100 1000
Frequency [MHz]
Amplitude AC rms [V]sinus