Use Instructions

General Information
20 / 77
CAUTION!
Service connection
Service connection is reserved for service tasks . The connected equipment must
comply with the standard EN 60950 in the currently valid version. Do not connect a
device to the service connection during surgery.
CAUTION!
Peripheral Devices
Additional peripheral equipment connected to interfaces of the medical monitor
has to meet the requirements of the following specifications in the respective cur-
rent valid version: IEC 60601-2-18 / EN 60601-2-18 for endoscopic devices and
IEC60601-1 / EN60601-1 for electrical medical devices. All configurations have to
comply with IEC60601-1 / EN60601-1 specifications. Whoever connects additional
equipment to signal output or signal input is considered the system configurator
and as such is responsible for complying with requirements of the standard
IEC60601-1 / EN60601-1.
CAUTION!
Mains Power Cable
Any power cables employed by the user that are not provided by the manufacturer
must meet the safety requirements of the national standards in the respective cur-
rent valid version.