Use Instructions

Safety Functions
54 / 77
8.3 Contamination
WARNING!
Contamination
Do not use device and/or accessories if signs of contamination are detected. Make
sure the device or/and accessories can no longer be operated until a qualified ser-
vice technician conducts the appropriate tests and repairs.
Error message "Contamination"
When fluid has penetrated the device via the insufflation tube connection, the dis-
play depicts
Contamination
and gives an acoustic signal. It is possible to conclude
any currently ongoing surgery with this device.
If insufflation is stopped subsequently, it can be restarted within 20minutes and
the surgery can be continued. A change of operating mode is no longer possible. If
insufflation is not restarted during 20minutes or if the device is turned off and on
again, further use of the device is not possible. This is to prevent cross-contamina-
tion.
Contamination
When you turn on an already contaminated device, a corresponding message ap-
pears on the display. The device can no longer be used. The contaminated device
has to be clearly marked as contaminated and sealed in two separate protective
layers of safety foil. Make sure the device cannot be used until a qualified service
technician conducts the appropriate tests and repairs.
8.4 Leakage Detection
If the device is not able to achieve the set pressure, a leakage notification is dis-
played.
Leakages can occur at port sides, port sealings or due to excessive use of a suction
device.