Electromagnetic compatibility
63 / 77
Emissions test Compliance Electromagnetic environment - guidance
RF emissions CISPR 11 Group 1
The device
PNEUMOCLEAR™
uses RF energy solely for its
internal function. Therefore, its RF emissions are very low
and are not likely to cause any interference in nearby elec-
tronic equipment.
RF emissions CISPR 11 Class B
The device
PNEUMOCLEAR™
is suitable for use in all estab-
lishments, other than public establishments and those
directly connected to the public mains network that sup-
plies buildings used for public purposes.
Harmonic emissions
IEC61000-3-2
Class A
Voltage fluctuations/ flicker
emissions IEC61000-3-3
Complies
11.3 Guidelines and Manufacturer's Statement - Electromagnetic
Interference Immunity
The device
PNEUMOCLEAR™
is intended for use in an electromagnetic environment
as described below. The user/operator of the device
PNEUMOCLEAR™
should make
sure the device is operated within such an environment.
Electromagnetic inter-
ference immunity tests
IEC 60601 test level Compliance levels Electromagnetic environment guidelines
Electrostatic discharges
(ESD) IEC61000-4-2
± 8kV contact
± 2, ± 4, ± 8, ± 15kV air
± 8kV contact
± 2, ± 4, ± 8, ± 15kV air
Floors should be made from wood or concrete or
covered with ceramic tiles. If the floor covering
consists of synthetic material, the relative humid-
ity should be at least 30%.
Electrical fast transi-
ents/bursts according to
IEC61000-4-4
± 2kV for power lines
Modulation100kHz
± 2kV for power lines
Modulation100kHz
The quality of the supply voltage should be the
same as the voltage of a typical business or hos-
pital environment.
Surges according to
IEC61000-4-5
±1kV line(s) to line(s)
±2kV line(s) to earth
±1kV line(s) to line(s)
±2kV line(s) to earth
The quality of the supply voltage should be the
same as the voltage of a typical business or hos-
pital environment.
Blackouts, brownouts,
and fluctuations of the
power supply according
to IEC61000-4-11
0 % UT; 0.5 cycle
At 0°, 45°, 90°, 135°,
180°, 225°, 270° and
315°
0 % UT; 0.5 cycle
At 0°, 45°, 90°, 135°,
180°, 225°, 270° and
315°
0 % UT; 1 cycle and
70 % UT; 25/30 cycles
Singlephase: at 0°
0 % UT; 1 cycle and
70 % UT; 25/30 cycles
Singlephase: at 0°
0 % UT; 250/300cycles 0 % UT; 250/300cycles
Supply frequency
(50/60 Hz) magnetic
field IEC61000-4-8
30A/m 30A/m
Magnetic fields of the mains power frequency
should comply with the typical values of business
and hospital environments.
Conducted RF
IEC61000-4-6
3Vrms
0.15MHz to 80MHz
6 Vrms in ISM bands
between 0.15 MHz and
80MHz
80 % AM at 1kHz
3Vrms
0.15MHz to 80MHz
6 Vrms in ISM bands
between 0.15 MHz and
80MHz
80 % AM at 1kHz
Transients RF
IEC61000-4-3
3 V/m 0.15 MHz to
80MHz
80%AM by 1kHz
3 V/m 0.15 MHz to
80MHz
80%AM by 1kHz