printed on them. Make sure the person has the same clothing on he or she was wearing when
measuring and calibrating. The weight scale is accurate to within 0.2 kg.
**Calibration result will be more accurate if there are two persons calibrating; one standing on the
platform and the other typing the weight.
Calibrating On-line Peripherals
When initially setting up your Polar BodyAge® System and each time the Polar BodyAge®
System is transported or every two months, it is recommended to calibrate the skinfold calipers,
the load platform (which calibrates both scale and biceps strength test) and the range of motion
unit to ensure accurate results.
Select System Settings > Calibrations. Select the Calibration Type from the drop-down box.
You will be prompted with instructions regarding how to calibrate the equipment. For further
instructions, see the following sections:
Calibrating the Skinfold Calipers
Calibrating the Load Platform
Calibrating the Range of Motion
Calibrating the Range of Motion
1. Select System Settings > Calibrations.
2. Select Range of Motion from the drop-down box. Click Calibrate.
120