4. If using the metal dowel provided with the system, place it lengthwise between the calipers as
shown. The measurement of the dowel is exactly 25.4 (that is the default value). If you are
measuring something else than the dowel, enter the result in the text box. Click Next.
5. Make sure the current calibration reading is between 14.0-17.5