• If the readings are not within 24.5mm and 26.0mm, re-calibrate and repeat.
SIT & REACH (Range of Motion)
Range of motion (ROM) handle is not retracting
• If the ROM handle is not retracting, the internal mechanism needs to be repaired by
Polar/HealthFirst service personnel.
• Remove the ROM unit from the sit & reach station using a 5/32” Allen wrench provided with your
system.
• Remove the two upper screws mounting the unit to the station.
• Unplug the ROM cable from the system.
• Remove the ROM unit.
Range of motion is inconsistent
• Re-calibrate ROM. See
Calibrating the Range of Motion for instructions.
• Enter into a client test and perform another sit and reach test.
• When the screen prompts to “Zero the System”, pull the handle out exactly 30 cm (11.8 inches)
from the footplate and press the Enter key.
• Return the handle slowly to the footplate and allow the TriFIT to take the reading.
• The measurement should be within 1 cm (0.4 inches) of 30cm (11.8 inches).
• If the measurement is not within 1 cm (0.4 inches) , re-calibrate and repeat.
Limited Polar International Warranty for Polar bodyage System
Polar Electro Group (Polar) warrants its equipment for one year from the date of shipment. This
warranty is limited and is in lieu of all other warranties expressed or implied. Within the warranty
period, Polar or its local representative will repair or replace, without charge, any parts, which
become defective due to material or workmanship.
Accessories, components and major sub-assemblies of Polar systems not manufactured by Polar
(i.e. Bike Ergometer, HP Computer, Tanita Body Composition Analyzer) will be warranted
according to the terms of the original manufacturer. This warranty shall not apply to any part or
parts of the system if it has been altered, repaired or misused through negligence; nor does this
warranty cover replacements or repairs necessitated by loss or damage resulting from any cause
81