User's Manual

TES-63 Source
User manual version 06/2021
Page 36/46
Polygon Physics SAS au capital social de 72 000 euros - 19, rue de Sassenage - 38600 Fontaine
SIREN 799 584 453 RCS Grenoble - TVA intracommunautaire FR 19 799584453 - APE 2899B
Can be used to decrease
voltages.
7.6 Getting state and setpoints with <get> <devicename><cr>
Command
Desciption
Comments
get id
Returns rack name.
get version
Returns the software version.
get rack_state
Returns the rack state.
0 = rack idle
1 = rack ready to operate
get hf_state
Returns the HF generator state.
0 = off
1 = on
get beam_state
Returns beam high voltage state.
0 = off
1 = on
get focus_state
Returns focus high voltage state.
0 = off
1 = on
get beampolar_state
Returns beam high voltage
polarity state.
0 = positive polarity
1 = negative polarity
get focuspolar_state
Returns focus high voltage
polarity state.
0 = positive polarity
1 = negative polarity
get hf
Returns HF power setpoints in
W.
get beam
Returns beam high voltage
setpoint in kV.
get imaxbeam
Returns the maximum current
setpoint delivered by the beam
power supply in microampere.
get focus
Returns focus high voltage
setpoint in kV.
get imaxfocus
Returns the maximum current
setpoint delivered by the focus
power supply in microampere.
get gasflow
Returns the gas valve opening
setpoint in V.
get network
Returns the IP address and
subnet mask.
get totaltime
Returns the total time of use of
the source in hours.
Ex.: the answer « 65.3 »
means the source has been
functioning 65.3h since
previous reset of totaltime.
get source
Returns all the states and
See list here after.