SIREN 799 584 453 RCS Grenoble - TVA intracommunautaire FR 19 799584453 - APE 2899B
meas hf
Returns the measurement
of the high frequency power
delivered by the HF
generator in W.
meas beam
Returns the measurement
of the beam voltage kV.
meas ibeam
Returns the measurement
of the current delivered by
the beam power supply in
µA.
meas focus
Returns the measurement
of the focus voltage kV.
meas ifocus
Returns the measurement
of the current delivered by
the focus power supply in
µA.
meas interlock
Returns a measurement of
the interlock state. If the
result is greater than or
equal to 4.9V, interlock is
closed (HF power and high
voltages are available).
meas source
Returns all the measured
states and setpoints. The
values should very close
(equal for states) to the
values returned by <get
source><cr>.
See list here after.
Description of the list returned by <meas source><cr>
This command returns a list whose elements are:
1.
hf_state
2.
beam_state
3.
focus_state
4.
beampolar_state
5.
focuspolar_state
6.
de_state
7.
state7 (Not connected)
8.
state8 (Not connected)
9.
HF