User guide

Model 530 Operating Instructions
841 Holt Road #1 Webster, NY 14580 Tel: 585-872-9350 Fax: 585-872-2638 sales@piecal.com http://www.piecal.com/
Practical Instrument Electronics, Inc. Copyright 2002. All rights reserved.
530-9002 2/26/04
8-8
Source/Power and Measure 2-Wire Transmitter Specifications:
Loop compliance voltage 24 Volts
Loop drive capability 1200 at 20 mA for entire battery life
Miscellaneous Open loop or out of compliance conditions are indicated by appropriate error display
Battery life in:
Source mode 18 hrs at 12mA typical (HART
®
disabled)
Power measure 10 hrs at 12mA typical
HART
®
protocol mode is a selectable option at turn on. HART
®
protocol mode places a
250 resistor in series with the output
Selectable EZ-Step(s) for Source Mode/2-Wire Transmitter Simulation:
In mA mode: 0.001, 0.010, 0.100, 1.000, 4.000(default), 8.000 mA
% of 4-20 mA mode: 0.01, 0.10, 1.00, 10.00, 25.00(default), 50.00 %
Read mA Specifications:
Voltage burden 2V at 20 mA
Overload/Current limit protection nominal 24 mA
Battery life Typical 40 Hours
2-Wire Transmitter Simulation Specifications:
Voltage burden 2V at 20 mA
Overload/Current limit protection nominal 24 mA
Loop voltage limits 2-60 VDC
Miscellaneous Open loop or out of compliance conditions are indicated by appropriate error display
Battery life 40 hour typical
Selectable EZ-Step(s) for Source Mode/2-Wire Transmitter Simulation:
In mA mode: 0.001, 0.010, 0.100, 1.000, 4.000(default), 8.000 mA
% of 4-20 mA mode: 0.01, 0.10, 1.00, 10.00, 25.00(default), 50.00 %
Voltage Read Specifications:
Range 0.00 to 60.00 VDC (with 2X over range)
Accuracy ± (0.1 % of Reading ±0.1 V)
Temperature effect ± 200 ppm/°C of Reading
Resolution 0.01 V
Input resistance 1 M
Battery life > 40 hour typical
Flashing indicator for over range
Warranty
Our equipment is guaranteed against defective material and workmanship (excluding batteries) for a period of three years from the
date of shipment. Claims under guarantee can be made by returning the equipment prepaid to our factory. The equipment will be
repaired, replaced or adjusted at our option. The liability of Practical Instrument Electronics (PIE) is restricted to that given under our
guarantee. No responsibility is accepted for damage, loss or other expense incurred through sale or use of our equipment. Under no
condition shall Practical Instrument Electronics, Inc. be liable for any special, incidental or consequential damage.