Section 4 - Maintenance and Parts 
Routine Maintenance 
Perform the following maintenance on a minimum of 3 month time cycle: 
  Remove dust from radiator by means of compressed air, in order to insure optimum 
cooler performance. 
  Check the filter screen for sediment, by removing the 15/16" acorn nut on the pump.  
Clean the filter or replace as necessary. 
 Empty and flush water reservoir yearly. Remove filler cap and fill reservoir with 
mixture of distilled water (4 gallons) and propylene glycol (1 gallon, RV type, pink 
color) solution. Do not use anti-freeze with stop leak additive, which will clog the filter 
screen and damage the pump. 
Parts 
Part Number  Product Description 
086813-004  C100 Base Unit (cooling unit only) 
106217-001  Fitting and Hose Kit (connect 1st heater) 
106218-001  Tee Kit (connect additional heaters) 
099076-001  Spare water pump 
120034-001  Anti-freeze, water conditioner 
Table 3 - C100 Parts and Kits 