Example:
Send TCP/IP command to 9816 module (via its connected socket) that returns decimal
Voltage data (of the raw temperature signal) for channels 1, 5, 9, and 13:
n11110
Response contains data for channels 13, 9, 5, and 1 (left to right):
"0.53013 0.541698 0.503633 0.000000"
In this example channels 13, 9, and 5 return normal temperature voltage signals in the
range of 0.5 to 0.6 volts. Note that channel 1 returns a value of 0.0 volts, indicating a
possible error in its temperature signal.