OWNER’S MANUAL Operation and Installation Prestige PM-D3 Washer Disinfector (PRFL0202) PRESTIGE MEDICAL P.O. Box 154, Clarendon Road Blackburn, Lancashire, BB1 9UG. England. Tel: Fax: E-mail: Web: (+)44 (0) 1254 682622 (+)44 (0) 1254 682606 sales@prestigemedical.co.uk www.prestigemedical.co.
Prestige PM-D3 Washer Disinfector Section 1 Section 1 ….
Prestige PM-D3 Washer Disinfector Section 1 Memorandum of Registration …. Note to Person Installing this Product This Memorandum of Registration MUST be completed by the person undertaking installation and witnessed by the owner of the product. Note to Owner of this Product Be sure that this Memorandum of Registration is completed by the person undertaking to install the product.
Prestige PM-D3 Washer Disinfector Section 1 Important Information …. ! IMPORTANT: Installation and Maintenance The loading door cannot be opened unless the product is connected to a mains power supply. For manual release, see page 10. Observe the Installation Instructions (page 23), the Maintenance Instructions (page 11), and the Servicing requirements outlined on page 27 of this manual. ! Product Identity The model and serial number are located on the silver label to the rear of the unit.
Prestige PM-D3 Washer Disinfector Section 1 Contents ….
Prestige PM-D3 Washer Disinfector Section 1 Introduction …. Intended Purpose and Performance Changing Language The intended purpose of the Prestige PMD3 Washer Disinfector is to clean, or clean and disinfect surgical instruments and associated products. This is achieved through the use of separate wash and disinfection stages which are controlled and monitored using sophisticated electronics to ensure optimum decontamination.
Prestige PM-D3 Washer Disinfector Section 1 Identifying Controls …. ! DOOR RELEASE IMPORTANT The door cannot be released unless the product is connected to a mains power supply. Press to open door Also used to select alternative programmes To manually open door, please refer to page 10. STATUS INDICATORS READY Unit will begin cycle when START button is pressed D3 PRE-WASH LCD DISPLAY Cold water pre-wash stage of cycle is in progress 1 WASH 2 Main wash stage is in progress.
Prestige PM-D3 Washer Disinfector Section 1 Operating Instructions …. Operating the PM-D3 Washer Disinfector 1. Apply power to the machine and press the DOOR RELEASE button to open the door. 2. Load the instruments to be cleaned into the Large Basket, Small Instrument Basket, Plate Rack, and Instrument Trays as required. Slide Large Basket into chamber. ! Do not overload the Large Basket so that water spray is obstructed.
Prestige PM-D3 Washer Disinfector Section 1 Operating Instructions (cont.) …. IF All Loaded Press START If Decono~Zyme has already been loaded then continue. If not, then return to step 3. Press START button once The Wash and Disinfection cycle will now commence and proceed automatically as detailed in “Operating Process” on page 25 of this manual.
Prestige PM-D3 Washer Disinfector Section 1 Changing Programmes …. Wash & Disinfect Function Changing Programmes Three separate programmes allow the Prestige PM-D3 to function as a washer disinfector, or as a washer only. The two wash/disinfect programmes enable a choice of disinfection times, whilst the wash only mode provides a 55°C final rinse stage to enable instruments to dry faster before being transferred to a separate disinfector or sterilizer.
Prestige PM-D3 Washer Disinfector Section 1 Emergency Door Release …. If your Prestige PM-D3 Washer Disinfector should fail to operate due to a fault with the supply to the product, the loading door will remain locked until power is restored. However, in the unlikely event that an internal fault causes a failure, the loading door must be released manually to gain access to the screws securing the outer cover.
Prestige PM-D3 Washer Disinfector Section 1 Routine Checks …. To ensure that your Prestige PM-D3 Washer Disinfector continues to operate within the design parameters it is recommended that the following checks are performed on a regular basis. Consulting the Trouble-Shooting guide on page 26 of this manual may prove useful when conducting the checks. Please note that completion of these checks is not a substitute for regular servicing and maintenance by a suitably qualified engineer.
Prestige PM-D3 Washer Disinfector Section 1 AWT Fluid Replenishment …. When full, the AWT Fluid reservoir contains sufficient anti-scaling fluid for around 40 wash and disinfect cycles. On completion of 25 cycles, the LCD DISPLAY will indicate that more fluid is required and the unit will not proceed with a new wash and disinfect cycle until you confirm that it has been added. To add AWT Fluid, the following procedure applies: 1. Apply Power to the Machine. Press DOOR RELEASE to open door. 2.
Prestige PM-D3 Washer Disinfector Section 1 Warranty …. PRODUCT WARRANTY To ensure continued efficiency and long operational life, your new Washer Disinfector is covered by a one year or 500 cycles (whichever occurs first) warranty which is arranged and administered by Prestige Medical. Details of the warranty are explained in this section.
Prestige PM-D3 Washer Disinfector Section 2 Section 2 ….
Prestige PM-D3 Washer Disinfector Section 2 Unpacking …. Unpacking Enclosures Check List Read this manual and familiarise yourself with all components and controls before attempting to install or operate the unit. Carefully unpack your Prestige PM-D3 Washer Disinfector and retain the carton and all packaging materials. If the unit needs to be returned to the supplier for any reason then the original carton and packaging materials should be re-used.
Prestige PM-D3 Washer Disinfector Section 2 Technical Specification ….
Prestige PM-D3 Washer Disinfector Section 2 Technical Specification (cont.) …. Required Services (cont): Drain: Discharge Orifice Type: Max. Flow to Drain: Max. Temp. of Effluent: 18mm outside dia. unthreaded hose tail 4 litres/min 85°C Ventilation: Max. Heat Output of Load: 85°C Chemicals: Size and Nature of Containers: AWT Fluid: 0.
Prestige PM-D3 Washer Disinfector Section 2 Chemical Data …. This document conforms with Regulation 6 of the Chemicals (Hazardous Information and Packaging for supply) Regulations 1994 (CHIP 2) - Second Edition Revision No: Date of Publication: Product: M105/01 21/01/02 AWT Liquid 1. IDENTIFICATION OF THE PRODUCT AND COMPANY 1.1 1.2 1.3 1.4 Trade Name: Type of product: Manufacturer/Supplier: Marketing Address: 1.
Prestige PM-D3 Washer Disinfector Section 2 Chemical Data (cont.) …. 8. EXPOSURE CONTROLS/PERSONAL PROTECTION 8.1 Personal protection: 8.2 Skin contact: 8.3 8.4 8.5 8.6 Eye contact: Inhalation: Inhalation (long term): Ingestion: Use with care, avoid eye contact and prolonged skin contact. Gloves and safety glasses recommended if available. External use only. Not for ingestion. Low risk: may degrease skin leading to dryness if excessive contact. Low risk: may cause temporary discomfort.
Prestige PM-D3 Washer Disinfector Section 2 Chemical Data (cont.) …. This document conforms with Regulation 6 of the Chemicals (Hazardous Information and Packaging for supply) Regulations 1994 (CHIP 2) - Second Edition Revision No: Date of Publication: Product: M104/01 21/01/02 Decono~Zyme Gel 1. IDENTIFICATION OF THE PRODUCT AND COMPANY 1.1 1.2 1.3 1.4 Trade Name: Type of product: Manufacturer/Supplier: Marketing Address: 1.
Prestige PM-D3 Washer Disinfector Section 2 Chemical Data (cont.) …. 8. EXPOSURE CONTROLS/PERSONAL PROTECTION 8.1 Personal protection: 8.2 Skin contact: 8.3 8.4 8.5 8.6 Eye contact: Inhalation: Inhalation (long term): Ingestion: Use with care, avoid eye contact and prolonged skin contact. Gloves and safety glasses recommended if available. External use only. Not for ingestion. Low risk: may degrease skin leading to dryness if excessive contact. Low risk: may cause temporary discomfort.
Prestige PM-D3 Washer Disinfector Section 2 Dimensions ….
Prestige PM-D3 Washer Disinfector Section 2 Installation …. Connecting the Prestige PM-D3 Washer Disinfector: 1. Unpack the machine and place it on a sturdy, level surface. The unit should be positioned within 1.83m of a mains supply (two supplies if the optional printer has been specified), a suitable drain, and a suitable cold mains water supply. 2. Plug the mains plug into a suitable switched outlet and switch on. 3.
Prestige PM-D3 Washer Disinfector Section 2 Setting Date and Time …. 1. Open loading door. 2. Press and hold DOOR RELEASE BUTTON. 3. After a few seconds the following screen will be displayed with a prompt flashing before the HOUR segment of display. Press and hold START button until correct HOUR is displayed. Set Clock & Prog TIME _00:00:40 4. Momentarily press START button 5. The following screen will now be displayed with prompt flashing between the HOUR and MINUTES segments of display.
Prestige PM-D3 Washer Disinfector Section 2 Operating Process …. small quantity of scale-inhibitor (AWT Fluid) is automatically injected into the rinse water.
Prestige PM-D3 Washer Disinfector Section 2 Troubleshooting Guide …. The following table will enable a non-technical user to diagnose basic problems and affect a cure that will enable the unit to resume operation as soon as possible. If the problem persists after performing these checks it will be necessary to contact your supplier.
Prestige PM-D3 Washer Disinfector Section 2 Service Record …. Medisafe recommend that your PM-D3 Washer Disinfector is serviced in accordance with the schedule contained within the PM-D3 Service Manual at intervals of 12 months (or 3 months if full HTM2030 compliance is required). This record must be completed by the Engineer conducting the Service and should be made available in the event of a warranty claim or breakdown repair. No user serviceable parts inside.
DECONTAMINATION CERTIFICATE ENSURE THAT ITEMS RETURNED TO YOUR SUPPLIER ARE DISINFECTED IN ACCORDANCE WITH THE APPROPRIATE HOSPITAL PROCEDURE FOR EQUIPMENT OF THIS TYPE This form MUST be completed and attached to the outside of the carton when returning the product to your supplier. Without this form being visible the courier and/or your supplier are at liberty to refuse handling of the product.