User's Manual
40
Chapter 2 Control Panel
Pneumatic Control Valves And Gauges
Air Cylinder Regulator
NOTE: You may also adjust the cylinder delay time through the Applicator 
Delay menu. See “Applicator Delay Menu” on page 195.
The air cylinder regulator (CYLINDER valve) is used to regulate the air to the 
applicator cylinder. The regulator setting affects how quickly the applicator 
pad will extend (apply stroke) and return (return stroke) during the apply 
cycle. This adjustment determines the force with which the applicator pad will 
contact the product. If set too high, the applicator pad could contact the 
product with enough force to cause damage. If set too low, the applicator pad 
may not contact the product. The air pressure delivered for the apply stroke 
and the return stroke is equal.
NOTE: Do not exceed 60 psi.
Monitor the setting of the air cylinder regulator using the PRESSURE gauge.
Air Jet Adjustment
The air jet adjustment (AIR JET) controls the air supply to the air jet tube. Air 
is forced out the air jet tube allowing the labels to properly transfer over the 
peel bar then to the applicator pad. If the adjustment is too low, the labels will 
not properly transfer from the printer to the applicator pad. If set too high, the 
label edge will be incorrectly positioned. The air jet is factory preset. Adjust 
the air jet pressure by turning the AIR JET control clockwise to increase the 
flow, counterclockwise to decrease the flow. See “Positioning The Air Jets” on 
page 52.
Vacuum Adjustment
NOTE: You may also adjust the vacuum delay time through the Applicator 
Delay Menu. See “Applicator Delay Menu” on page 195.
The vacuum adjustment (VACUUM) controls the amount of air flow through 
the vacuum generator, thus determining the amount of vacuum holding the 
label onto the applicator pad. A weak vacuum will cause labels to fall off of the 
applicator pad prematurely. A vacuum that is too strong, however, can cause 
difficulty when transferring the label onto the applicator pad and the product. 
Vacuum is increased by rotating the valve clockwise and decreased by 
rotating the valve counterclockwise.
NOTE: To get a vacuum on every label, every hole on the applicator pad 
must be covered by the label surface.
Monitor the setting of the vacuum adjustment using the VACUUM gauge. 
The minimum setting is 20 inches hg.










