Specifications
2.39
Applications Manual
© 2003-2005 Prism Media Products Ltd
Revision 1.11Prism Sound dScope Series III
Specification
No precise specification is provided. As a guide, separation should generally be better than 85dB at
997Hz.
4.2.1.7 Frequency Response
dScope Configuration:
Frequency Response Track 19.dsc
Track Number:
19
Corresponding AP Test File:
2_frqswp.at2
.ac3 Test File:
2fstplr.ac3
Duration:
60 sec
Test Signal Description:
20Hz to 20kHz sweep at –20dBFS
Description
This test plots the frequency response of the
EUT
at –20
dBFS
.
Method
Click the
[OK]
button to start the test, and then play Track 19.
The dScope senses the 60 logarithmically-spaced frequencies on the test track, measures the
amplitude at each (using a band-pass filter to exclude noise), and produces a plot of the results for
both channels. Note that the output scale is referenced to the nominal –20dBFS level of the test