Specifications
2.46
Applications Manual
© 2003-2005 Prism Media Products Ltd
Revision 1.11Prism Sound dScope Series III
Specification
Dolby Laboratories does not provide a precise specification, however the red box on the plot shows
the minimum eye region specified by the IEC60958 standard, into which the eye pattern should not
transgress if reliable reception is to be guaranteed.
4.2.2.2 Jitter Spectrum
dScope Configuration:
Jitter Spectrum Track 19.dsc
Track Number:
(any)
Corresponding AP Test File:
Digjitn.at2
.ac3 Test File:
(any)
Duration:
Test Signal Description:
Description
This test measures the worst-case peak
jitter
of the
EUT
's digital output carrier, and also displays the
spectrum of the jitter. Excessive jitter can lead to the output data being irrecoverable by a receiving
device. Even moderate jitter can result in degraded audio performance in a subsequent D/A
converter if its design is inadequate in rejecting the incoming jitter from its sampling clock.
Method
Play Track 19 and then click the
[OK]
button to start the test.