Specifications

2.76
Applications Manual
© 2003-2005 Prism Media Products Ltd
Revision 1.11Prism Sound dScope Series III
5.2.1.14 Level Trim Controls
dScope Configuration:
Level Trim Controls Track 30.dsc
Track Number:
30
Corresponding AP Test File:
N/A
Duration:
30 sec
Test Signal Description:
997Hz, -20dBFS
Channels:
L, C, R, Ls, Rs
Description
This test verifies the operation of level trim and/or balance controls, if provided.
Method
Set all level trim/balance controls to their nominal positions.
The script will go through each channel in turn, and ask whether the channel has trim/balance
controls. If it has, then the test will be performed for the channel.
For each channel with level/trim controls, play track 30 and then click the
[OK]
button to start the test.
The nominal level of the channel wil be recorded and a Sweep performed. Increment the trim for the
channel by one step, then restart track 30 and click
[OK]
again. The Sweep will be repeated at this
new level. This process will be repeated again for the minimum and maximum level of the trim control.