Specifications

Prism Media Products Limited AD-2 Dual-rate A/D converter
Operation Manual
Issue 0.2 © Prism Media Products Limited
11 August 1998 Page 51 of 58
There are five fuses internal to the unit, located close to the main heatsink. If these
fuses have blown, a fault has occurred and the unit should be returned to the
manufacturer or agent for service. The fuse types and applications are:
F3 - Analogue 5V supply
F4, F5 - Digital 5V and 3V3 supplies
Type: 2AT Wickman TR5 subminiature type no. 19372K/2A
F1, F2 - Analogue input supplies
Type: 0.5AT Wickman TR5 subminiature type no. 19372K/500mA