INDEX 1. Instrument overview 2. How to use the instrument 2:1 Description 2:2 Display 2:3 On/Off & push buttons 2:4 Settings • Set date / time • Change number of digits • Change limit value • Change program mode 2:5 Measuring crankshaft deflection 2:6 Print documents 3. 4. 5. 6. 7.
2. HOW TO USE THE INSTRUMENT 2:1 DESCRIPTION The importance of accurate crankshaft deflection measurement can´t be enough emphasized. The DI-5C series of instruments are designed to be both accurate, fast and user friendly. The complete instrument set is stored in a customized carrying case which includes Main instrument, Battery Charger, Transducer, Connecting cable, Extension bars, Measuring tip, Program Manual and User Manual.
OK < To START the instrument, press OK. The OK button is used to accept throughout the step by step procedure. To SWITCH OFF the instrument, press and hold OK for 3 seconds or , when in date/time mode, press < twice to reach the menu alternative SWITCH OFF INSTRUMENT? Accept by pressing OK. Used to change value or position > Used to change value or position > 2:3 ON/OFF, PUSH BUTTONS This button allows you to step backwards in the menu if you should enter and accept OK by mistake.
2:4.4 Change program mode When selecting STANDARD MODE by OK, the program (menu) will follow the Flow Scheme on the last page. If MEASUREMENT ONLY is selected the display will show current date. By pressing OK the instrument will remain in measurement only mode. NOTE ! In this mode the panel buttons operate in the following way; OK = ON/OFF < = RESET > = ZERO SET ^ = BACKLIGHT on/off.
p) q) r) s) t) u) v) Set the instrument to ZERO by pressing OK. Select position ‘A’ for clockwise rotation and ‘E’ if anti clockwise. With instrument reading 0,000 mm, rotate the crankshaft to next position and accept the reading by pressing OK. Continue through the remaining positions of rotation. When all positions are complete, the instrument will display the next cylinder in line order, press OK to accept.
2:8 FIND STORED DOCUMENTS To find stored documents you should follow the procedure below; • • • • • • Start the instrument by pressing OK, accept date/time with OK The display will ask NEW DOCUMENT? Pressing the right > arrow the display will move to FIND STORED DOCUMENT, press OK The display will now ask you to choose DOCUMENT NR. Using the right > arrow key, you move to the required document, then press OK.
2:11 DELETE DOCUMENTS This menu provides two options; delete a single document or, delete all documents. Delete single document: Follow section 2:8 above until DELETE DOC is displayed, press OK. The display will read ARE YOU SURE, NO ? If ‘NO’ press OK, the display will move to FIND OTHER DOCUMENTS. If arrow right > is pressed the display shows ARE YOU SURE, YES ? If ‘YES’, press OK. Display will now show DOCUMENT DELETED and return to FIND STORED DOCUMENTS.
5. CALIBRATION OF INSTRUMENT Calibration of instrument is neccessary after change of transducer, done at Prisma Tibro, Sweden – send an email to support@prismatibro.se 6.
CHANGE LIMIT VALUE CHANGE PROGRAM MODE ENGINE NR? 01 MAIN ENGINE B AUXILARY ENGINE COLD ENGINE OVALITY MEASUREMENT WARM ENGINE CRANKSHAFT MEASUREMENT DOCUMENT NR: 00000000 FIND STORED DOCUMENTS? DATE FORMAT EU US DIESEL GENERATOR ENABLE 4 3 DIGIT MODE YEAR, MONTH, DATE, DAY, TIME MEMORY IS EMPTY NEW DOCUMENT NR: 00000000 XX OF 53 DOCS STORED ... USER DEFINED MAX LIMIT +0.500 MIN LIMIT -0.500 ARE YOU SURE NO? DELETE ALL DOCUMENTS? ALL DOCS DELETED WAIT...
ANTI-CLOCKWISE MEASUREMENT POS: X SAVED POSITION A 0.001 ZERO? NEXT CYLINDER ZERO NOT WANTED RESET NOT WANTED RESET? D MEASUREMENT DONE CURRENT CYL: NR 01 CYLINDER IN TOTAL 24 CLOCKWISE MEASUREMENT WATER TEMP 80 LUB OIL TEMP 70 AMBIENT TEMP 40 MEASUREMENT DONE RETURN TO RESET WAIT READ DOC DONE READ DOC DONE EDIT DOC? 00000000 ARE YOU SURE NO? ME01 CYL01 A X.