Caution: The control solution ranges are 
located on the back of the Prodigy® test strip 
vial. They are not recommended target ranges 
for your blood glucose.
Important Information:
• Use only Prodigy® control solutions with your
  Prodigy® meter.
• Check the expiration date on the bottle. 
  Do not use if expired.
• Use within a period of 90 days from the date
  that you rst open it.  Record the 
 discard date on the control solution bottle 
  when you rst open it to serve as a reminder 
  to discard after 90 days.
• For in vitro diagnostic use.
Why Perform a Control Solution Test:
• To ensure that your meter and test strips are
  working properly together.
• To allow you to practice testing without using
  your own blood.
When to use:
• Once a week (to ensure that you continue to
  have accurate results).
31
• When you begin using a new vial of test strips
• When test strips have been exposed to extreme
  environmental conditions.
• If you drop the meter.
• If you change the battery in the meter.
Storage and Handling:
• Store the control solution tightly closed at 
  temperatures below 86°F (30°C).  
• Do not refrigerate. Do not store in your car.
Composition:
  Glucose Nonreactive Ingredients 
  (% w/v)  (% w/v)
Low Solution  0.05  99.95
High Solution 0.19  99.81