Performing a Control Solution Test
Start with the meter off.
STEP 1: Wash your Hands
Wash your hands with mild soap and water 
before performing any test. Be sure to dry 
them thoroughly. 
STEP 2: Insert Test Strip
Insert a test strip with the contact bar end 
entering into the test strip slot rst. Push the 
test strip as far as it will go without bending it.  
The meter turns on automatically.
STEP 3: Mark as a Control Solution Test
After the “ ” symbol appears on the display, 
press the “M” button and a “ ” symbol appears 
on the display. With the “ ” symbol on the 
display, the meter will not store your test result 
in the memory. If you decide not to perform 
a control solution test, press the “M” button 
again and the “ ” symbol will disappear.
33
Important: Be sure that you are in control 
solution mode so that the test result will not be 
stored in the meter memory.
STEP 4: Apply Control Solution
1. Check the expiration and discard dates on
  your control solution and test strip vials.
2. Shake the control solution bottle well, then
  remove cap.
3. Squeeze the bottle and discard the rst drop
  then wipe the dispenser tip with clean tissue
  paper or cotton.
4. Squeeze the bottle again to
  get a second drop onto a
  clean, non-absorbent surface
  or on your clean ngertip.
  See Figure 18.
5. Bring the tip of the test strip 
  to touch the drop of solution
  until the meter beeps.
Figure 18