 Probable Cause Actions
Error in performing  Read the instructions
the test.  thoroughly and repeat
    the test again.
Control solution bottle  Shake the control
not shaken well.  solution bottle vigorously 
    and repeat the test again.
Expired or  Check the expiration
contaminated  date and the discard date 
control solution.  of the control solution.
Control solution that  Control solution 
is too warm or too cold. 
should come
    to room temperature 
    (less than 86°F/30°C) 
    before testing.
Test strip deterioration. Please repeat the test 
    with a new test strip.
  v The control solution test is out of 
 range.
62
 Probable Cause Actions
Meter malfunction.  Please contact 
    Customer Care at 
  1.800.243.2636 for 
  assistance.
  v The control solution test is out of 
 range.