Owner`s manual
PRODIGY
AutoCode® Blood Glucose Meter PRODIGY
AutoCode® Blood Glucose Meter
Owner’s Manual Owner’s Manual
70
Accuracy of Meter Measurements
FAQ: Can two (2) blood glucose meters produce 
different readings that are from the same blood 
sample? Does this make one reading more 
accurate than the other?
Answer: Yes, different meters can produce
  different readings using the same blood
 sample. 
  No, this does not necessarily mean that
  one result is more accurate than the
 other.
The reasons for this are as follows:
  1. The FDA recognized standard for glucose
    meter accuracy is ISO 15197: In Vitro
    Diagnostic Test Systems Requirements 
    for Blood Glucose Monitoring Systems for
    Self Testing in Managing Diabetes Mellitus.
    This standard sets accuracy requirements
    that manufacturers must meet before being 
    approved by the FDA to market their
 glucose meters.  
2. The accuracy requirements for the blood
  glucose monitoring system—meters and test
  strips—as stated in ISO 15197 are:
 Ninety-ve percent (95%) of the individual 
glucose results shall fall within ±15 mg/
dL (0.83 mmol/L) of the results of the 
manufacturer’s measurement procedure 
at glucose concentrations < 75 mg/dL 
(<4.2 mmol/L) and within ± 20% at glucose 
concentrations ≥ 75 mg/dL ( ≥ 4.2 mmol/L).
How does this answer the FAQ
The accuracy of the system is determined by a 
clinical study using 100 blood samples at interval 
glucose concentrations ranging from < 50 mg/
dL to over 400 mg/dL measured using both 
the manufacturer’s meter and a clinical blood 
glucose analyzer for comparison (reference 
measurement). 
Example 
I
f the manufacturer’s reference measurement is 70 
mg/dL this means that the meter measurements 
are considered accurate if they fall within ± 15 mg/
dL of 70 mg/dL (from 55 mg/dL to 85 mg/dL).
71










