16
•  Only apply a blood sample or a control 
solution sample to the test strip’s absorbent 
channel. Applying other substances to the 
test strip’s absorbent channel will cause 
inaccurate results. 
•  Record the discard date on the vial label 
when you rst open it. Discard remaining 
test strips 90 days after the rst opening 
date. 
•  Do not use test strips beyond the expiration 
date printed on the package.
•  After removing the test strip from the vial, 
insert it directly into the meter. Do not lay 
the test strip on any other surface.
Warning: Keep the test strip vial away from 
children; the vial cap and the test strips can 
be a potential choking hazard. Never chew or 
swallow a test strip. If this occurs, please seek 
medical assistance immediately.
17
 Introduction to Prodigy® Meters
Intended Use
Prodigy
®
 blood glucose monitoring systems 
are intended for use outside the body (in vitro 
diagnostic use only) and should only be used for 
blood glucose (blood sugar) testing with fresh 
capillary whole blood samples.
These systems are intended for use in the home 
and in clinical settings and should not be used 
for the diagnosis of diabetes or for the testing 
of newborns. Consult your healthcare provider if 
unusual readings occur.
Test Principle
Blood glucose is measured by an electrical 
current that is produced when a blood sample 
mixes with the reagent (special chemicals) of the 
test strip. The electrical current changes with the 
amount of glucose in the blood sample. Prodigy
®
meters measure the strength of the electrical 
current, calculate your blood glucose level 
and then display your result in either mg/dL or 
mmol/L.