Owner`s manual
Prodigy Diabetes Care, LLC  www.prodigymeter.com
PRODIGY
®
 Blood Glucose Meters   Owner’s Manual
20
21
Prodigy Diabetes Care, LLC  1.800.243.2636
PRODIGY
®
 Blood Glucose Meters   Owner’s Manual
• Sites other Than Your Fingertip 
Please refer to the “About Alternate Site Testing (AST)” Section. 
Please consult your healthcare professional before obtaining 
bloodfromsitesotherthanyourngertip.
STEP 3: Obtain a Blood Sample
Do not smear the blood sample.  
See Figure 13. To obtain the most 
accurateresults,wipeofftherst
drop of blood and gently squeeze another drop of blood. 
See Figure 14.
STEP 4: Remove the Lancet
Remove the cap of the lancing device by twisting it off. Place the 
protective cap back on the exposed tip of 
the lancet by pushing the lancet into it. See 
Figure 15. 
*Always use caution when removing the 
lancet. Discard the lancet according to 
your local safety regulations.
Warning:Therstdropofbloodusually
containstissueuidandserum,which
may affect the test result. It should be 
discarded.
Intended Use
Prodigy
®
 blood glucose monitoring systems are intended for use 
outside the body (in vitro diagnostic use only) and should only be used 
for blood glucose (blood sugar) testing with fresh capillary whole blood 
samples.
These systems are intended for use in the home and in clinical settings 
and should not be used for the diagnosis of diabetes or for the testing 
of newborns.
Test Principle
Blood glucose is measured by an electric current that is produced 
when a blood sample mixes with the reagent (special chemicals) of the 
test strip. The electrical current changes with the amount of glucose 
in the blood sample. Prodigy
®
 meters measure the strength of the 
electrical current, calculates your blood glucose level and then displays 
your result in either mg/dL or mmol/L.
Prodigy
®
 meters, test strips and control solutions have been designed, 
tested and proven to work together as a system to produce accurate 
blood glucose test results.
Important: Use only Prodigy
®
 control solutions and test strips with 
your Prodigy
®
 meter. Using other test strips and control solutions 
with this meter can produce inaccurate results.
Figure 14
Figure 13
Figure 15
Introduction to Prodigy
®
 Meters