Owner`s manual
Prodigy Diabetes Care, LLC  www.prodigymeter.com
PRODIGY
®
 Blood Glucose Meters   Owner’s Manual
76
77
Prodigy Diabetes Care, LLC  1.800.243.2636
PRODIGY
®
 Blood Glucose Meters   Owner’s Manual
 vThe control solution test is out of range.
  Probable Cause  Action
  Error in performing
  the test.
  Control solution bottle
  not shaken well.
  Expired or contaminated
  control solution.
  Control solution that is
  too warm or too cold.
  Test strip deterioration.
  Meter malfunction.
Read the instructions 
thoroughly and repeat 
the test again.
Shake the control solution bottle 
vigorously and repeat 
the test again.
Check the expiration date or 
the discard date of the control 
solution.
Control Solution, meter, 
and test strips should come 
to room temperature 
(68–77°F/20–25°C) 
before testing.
Please repeat the test with a 
new test strip.
Please contact Customer Care 
at 1.800.243.2636 
for assistance.
Comparing Meter and Laboratory Results
The test results you obtain from your meter may differ somewhat from 
your laboratory results due to normal variation. Meter results can be 
affected by factors and conditions that do not affect laboratory results 
in the same way (see test strip package insert for typical accuracy 
and precision data as well as important information on limitations). 
To make an accurate comparison between meter and laboratory 
results, follow the guidelines below.
Before you go to the laboratory:
• Perform a control solution test to make sure that the meter is   
working properly.
• It is strongly recommended to fast for at least eight (8) hours   
before doing comparison tests.
• Take your meter with you to the laboratory.
While at the laboratory:
• Make sure that the samples for both tests (the meter test and the 
laboratory test) are taken and tested within 15 minutes of each 
other.
• Wash your hands before obtaining a blood sample.
• Never use your meter with blood that has been collected in a gray 
top test tube.
• Use fresh capillary blood only.
You may still have a variation from the result because blood 
glucoselevelscanchangesignicantlyovershortperiods,
especially if you have recently eaten, exercised, taken medication or 
experienced stress.
1
 In addition, if you have eaten recently, the blood 
glucoselevelfromangerstickcanbeupto70mg/dL(3.9mmol/L)
Information About your Prodigy
®
 Meter