Owner`s manual
Prodigy Diabetes Care, LLC  www.prodigymeter.com
PRODIGY
®
 Blood Glucose Meters   Owner’s Manual
80
81
Prodigy Diabetes Care, LLC  1.800.243.2636
PRODIGY
®
 Blood Glucose Meters   Owner’s Manual
If your test result is lower than 70 mg/dL or “Lo” 
appears on your meter screen, this could be a result 
of hypoglycemia. Although this result could be due to 
atestingerror,itisbesttotreatrstandretestlater.
Contact your healthcare professional immediately.  
See Figure 99.
High Blood Glucose Test Results
If your test result is higher than 180 mg/dL, this could be a result of 
hyperglycemia. If you are uncertain about this test result, retest. If the 
result continues to be higher than 180 mg/dL, contact your healthcare 
professional immediately.
If “Hi” appears on your meter screen, this could be a result 
of severe hyperglycemia (results higher than 600 mg/dL). 
Retest your blood glucose. If “Hi” appears again, contact 
your healthcare professional immediately. See Figure 100.
Repeated Unexpected Test Results
If you continue to get unexpected test results, check your  
blood glucose monitoring system by performing a control solution test. 
Refer to “Performing a Control Solution Test” section in this manaul.
Unusual Red Blood Cell Count
A hematocrit range (percentage of red blood cells in the blood) that is 
extremely high (above 55%) or low (below 30%) can also cause false 
readings.
Figure 99
Figure 100
Accuracy of Meter Measurements
FAQ: Can two (2) blood glucose meters produce different readings 
that are from the same blood sample? Does this make one reading 
more accurate than the other?
Answer:  Yes, different meters can produce different readings using 
  the same blood sample. 
  No, this does not necessarily mean that one result is more 
  accurate than the other.
The reasons for this are as follows:
  1. The FDA recognized standard for glucose meter accuracy is 
    ISO 15197: In Vitro Diagnostic Test Systems—Requirements 
    for Blood Glucose Monitoring Systems for Self Testing 
    in Managing Diabetes Mellitus. This standard sets accuracy 
    requirements that manufacturers must meet before being 
    approved by the FDA to market their glucose meters. 
  2. The accuracy requirements for the blood glucose monitoring 
    system—meters and test strips—as stated in ISO 15197 
 is this:
Ninety-vepercent(95%)oftheindividualglucoseresults
shall fall within ±15 mg/dL (0.83 mmol/L) of the results 
of the manufacturer’s measurement procedure at glucose 
concentrations < 75 mg/dL (<4.2 mmol/L) and within ± 20% at 
glucoseconcentrations≥75mg/dL(≥4.2mmol/L).
How does this answer the FAQ?
The accuracy of the system is determined by a clinical study using 
100 blood samples at interval glucose concentrations ranging from 










