< 50 mg/dL to over 400 mg/dL measured using both the 
manufacturer’s meter and a clinical blood glucose analyzer for 
comparison (reference measurement). 
Example 1
If the manufacturer’s reference measurement is 70 mg/dL this 
means that the meter measurements are considered accurate if they 
fall within ± 15 mg/dL of 70 mg/dL (from of 55 mg/dL to 85 mg/dL).
If meter A produces a measurement of 60 mg/dL and meter B 
measures 80 mg/dL, then both are in the acceptable range and meet 
the accuracy requirement.
Each manufacturer has met the System Accuracy requirement as 
part of its testing to be cleared for marketing. In the System Accuracy, 
testing a distribution of blood glucose readings will result at the 
various glucose concentration levels for one manufacturer’s system. 
This is the same for other manufacturer’s systems. Because of the 
range of acceptable readings for the accuracy requirement, identical 
readings from two different meters on the same blood sample may 
not happen.
Symbols Information
    Do not re-use. Single use only.
    Consult Operating Instructions.
    Keep away from sunlight.
    Keep dry.
    Temperature limitation.
    Use by.
    Date of manufacture.
    Batch code.
    Catalog number.
    Serial number.
  Control.
LOT
REF
SN
CONTROL
 Symbol  Referent