MT-1217 3-3/4 Autorange Digital Multimeter User’s Manual st , 1 Edition 2013 ©2013 Copyright by Prokit’s Industries Co., Ltd.
Contents General-------------------------------------------------------------------------1 Open-Package Inspection-------------------------------------------------2 Safety Note -------------------------------------------------------------------2 Instrument Panel& Button Function Description----------------------5 Other Functions---------------------------------------------------------------6 Property-------------------------------------------------------------------------6 Instrument Maintenance------------
Open-package Inspection Open the package box and take out the meter, check carefully if the following accessories are absent or damaged. If there is any absence or damage, please contact the distributor immediately. Digital multimeter 1 PC User’s manual 1 copy Test leads 1 pair Temperature Probe (K-Thermocouple) 1 PC Test socket 1 pc Safety Note The design of this meter is accordance with IEC1010 clause (the safety standard issued by International Electro technical Commission).
5. Don’t use other unconfirmed or disapproved battery to replace the battery inside the meter. Only the same model or same electrical specification battery can be replaced. Before the replacement, the test lead must be removed from the measuring point to ensure there is no longer any signal at the input terminal. 6. When the electrical measurement is made, never let your body get in touch with the ground directly, and don’t touch uncovered metal terminal, output port, lead clamp and etc.
12. When “ ” is shown on LCD display, please replace battery immediately to ensure the measurement precision. 13. It is not allowed to insert the test lead into the current terminal to measure voltage! 14. Please don’t change the circuits of the meter freely for fear that the meter is damaged and the safety be endangered. 15.
3.3 HOLD: date hold 3.4 REL △(Relative Value Measurement): The relative value measurement of Capacity function could be conducted by pressing this button 3.5 Select(Function Switch): Press this button, the function ; could switch between resistance, capacity and AC/DC current。 3.6 LCD backlight button 4. Knob Switch: It could be used to change the measurement function and range. 5. Input Terminals 5.1. 10A “+”input terminal. 5.2. mA “+” input, Temperature “+” Input, Capacity and HFE input terminal 5.3.
1-3 Measuring Method: Dual integral A/D converter 1-4 Sampling Rage: Approx. 3 times / sec. 1-5 Over Range Indication: Display “OL” ” symbol appears; 1-6 Low Battery Indication: “ 1-7 Operation Environment: (0~40)˚C Relative Humidity:<80% 1-8 Storage Environment: (0~50)˚C Relative Humidity:<80% 1-9 Power: a 9V battery (NEDA1604/6F22 or equivalent) 1-10Dimension (size): 182×90×46mm 1-11 Weight: Approx.
Range Accuracy Resolution ±(0.5%+4d) 400mV 100uV 4V 1mV 40V 10mV 400V 100mV 1000V ±(1.0%+6d) 1V Input Impedance: 10MΩ . Overload Protection: 1000V DC or AC peak value 2-2-2. ACV A) Insert black test lead into the hole of “COM” and red test lead “ into “ “ setting , the initial state of B) Rotate function switch to “ the meter is in automatic range status, which shows “AUTO” symbol C) press “Range” button can select the AC measurement range D) Make the test lead contacted to the testing point.
Input Impedance: 10MΩ . Overload Protection: 1000V DC or 750 AC peak value. Frequency Response: (50~200) Hz Display: Average value response (RMS of sine wave). 2-2-3. DCA A) Insert the black test lead into the “COM” input terminal and red test lead into the “mA “ input terminal. (Max 400mA), or “10A “input terminal (Max 10A). B) Rotate function switch to Current setting. The initial state of the meter is in automatic range status, which shows “DC” symbol.
2-2-4. ACA A) Insert the black test lead into the “COM” input terminal and red test lead into the “mA “input terminal. (Max 400mA), or “10A” input terminal (Max 10A). B) Rotate function switch to Current setting. Press “SELECT” button to select the AC measurement mode. Then connect the test lead to the tested circuit in serial, the tested current value and the current polarity of the point where the red test lead is contacted will be displayed on the screen simultaneously. Caution: 1.
2-2-5. Resistance (Ω) A) Insert the black test lead into “COM” terminal and red test lead into “ “terminal. B) Rotate the Range to “Ω” setting. the initial state of the meter is in automatic range status, which shows “AUTO” symbol C) Connect the two test leads to the tested resistor. D) When measuring the low resistance, please short-circuit the test leads at first to test the wire resistance, and then deduct it from the actual resistance. Caution: 1.
WARNING: DO NOT input any voltage at resistance range for safety! 2-2-6.Diode and Continuity Test A) Insert the black test lead to “COM” terminal and the red test lead to “ ” terminal. (The polarity of red test lead is“+”). B) Rotate the Range to “ “setting. Press “SELECT” button to select the Diode measurement mode. C) Forward Measurement: Connect the red test lead to the diode positive polarity and the black test lead to the diode negative polarity.
• Overload Protection: 250V DC or AC Peak Value. WARNING: DO NOT input any voltage at this range for safety! 2-2-7. Capacity (F) A) Rotate function switch to “ “setting. B) Or rotate the Range to “Ω“setting. Press “SELECT” button to select the capacity measurement mode. C) Insert the black test lead to “COM” terminal and red test lead to “mA“ terminal; D) Connect the tested capacity by the test leads to “COM”, “mA “ input terminal, the screen will show capacitance parameter.
2-2-8. Frequency (Hz) A) Test volt frequency, connect test leads and shielded cable B) ” terminal. respectively to “COM” terminal and “ Rotate function switch to “ “ gear. Connect test leads and the cable to the signal source or the tested load. The tested signal will show on the screen. Caution: 1. When inputting AC RMS over 10V, it could show reading, but excess vibration may appear. 2. It is recommended to test weak signals by shielded cable under noisy circumstances. 3.
Caution: 1. When the input terminal is open-circuit, it will display the “OL” symbol on the screen. 2. Don’t change the temperature probe at random, or the value accuracy could not be guaranteed. 3. Don’t measure voltage at temperature range. Range Accuracy Resolution (-20~1000)˚C <400˚C±(1.0%+5d) 1˚C ≥400˚C±(1.5%+15d) • Sensor: K Type Thermocouple (Nickel-chromium--nickel silicon) (banana plug). WARNING: DO NOT input any voltage at this range for safety! 2-2-10.
Instrument Maintenance This is a precision instrument and the user shall not modify the electric circuit as well. 1. Keep the instrument away from water, dust and shock. 2. Do not store and operate the meter under the condition of high temperature, high humidity, combustible, explosive and strong magnetic place. 3. Wipe the case with a damp cloth and detergent; do not use abrasives and alcohol. 4. If the instrument is not operated for a long time, please take out the battery to avoid leakage. 5.
Fault Elimination If the instrument could not work properly, please try the following tips to solve some general problems. If the problems still exist, please contact the maintenance center or the distributor. Solution Fault Turn on power. No Display Or Replace battery symbol appearance No current input Error value Replace battery. Replace fuse Replace battery This Instruction is subjected to change without any further notice.
目 錄 一.概述----------------------------------------------------------------------------------------18 二.開箱檢查---------------------------------------------------------------------------------18 三.安全注意事項---------------------------------------------------------------------------19 四.儀錶面板及按鍵功能說明-----------------------------------------------------------20 五.其他功能--------------------------------------------------------------------------------- 21 六.
3. 4. 5. 6. 7. 8. 9. 10. 11. 12. 13. 14. 15.
四、儀錶面板及按鍵功能說明 1. 儀錶型號 2. 液晶顯示器:顯示儀錶測量的數值及單位 3. 功能按鍵說明 3.1. Hz/% (頻率/占空比):交流電壓位, 觸發此鍵, 可進行頻率/占 空比測量 3.2. Range: 量程選擇, 可手動選擇測 量的範圍以提高測量速度. 3.3. HOLD︰測量數據保持. 3.4. REL △(相對值測量):在電容檔 觸發此鍵, 可進行相對值測量模式. 3.5. SELECT: 選擇電阻, 電容, 二極管, 通 斷測量功能及電流檔 DC/AC 切換 3.6. LCD 顯示背光按鍵, 觸發此鍵控制開, 閉背光. 4. 旋鈕開關:用於改變測量功能及量程。 5. 輸入插孔 5.1. 10A“+"輸入插孔 5.2. mA“+"電流輸入, 溫度“+"輸入, 電容輸入以及電晶體輸入插孔 5.3. 電壓、二極體、電阻、頻率、蜂鳴器輸入插座 5.4.
六、特性 1.一般特性 1-1. 顯示方式:液晶顯示 1-2. 最大顯示:3999(3 3/4 位元)自動極性和單位顯示 1-3. 測量方式:雙積分式 A/D 轉換 1-4. 轉換速率:約每秒 3 次 1-5. 過量程顯示:顯示“OL" "符號出現; 1-6. 低電壓顯示:“ 1-7. 工作環境: (0~40)℃,相對濕度<80% 1-8. 相對溫度: (0~50)℃,相對濕度<80% 1-9. 電源:一隻 9V 電池(NEDA1604/6F22 或同等型號) 1-10. 體積(尺寸) :182×90×46mm (長×寬×高) 1-11. 重量:約 320 g(不包括 9V 電池) ; 1-12. 附件:說明書一本,10A 錶棒一付,K 型熱電偶一支, 測試座一個. 2.技術特性 2-1. 準確度:±(a%×讀數+字數),保證準確度環境溫度:(23±5)°C,相 對濕度<75%,校準保證期從出廠日起為一年。 2-2.
2-3-1.直流電壓(DCV) "檔 A). 將旋鈕開關轉至“ B). 儀錶初始為自動量程狀態,顯示“AUTO"符號 C.) 將測試錶棒接觸測試點,紅錶棒所接的該點電壓與極性顯示在螢幕 上。 注意: 1. 測量電壓切勿超過 1000V,超過則有損壞儀錶電路的危險。 2. 當測量高電壓電路時,千萬注意避免人體觸及高壓電路,注意人身安全。 量程 準確度 分辨力 400mV 100uV ±(0.5%+4d) 4V 1mV 40V 10mV 400V 100mV 1000V ±(1.0%+6d) 1V ‧輸入阻抗:10MΩ ‧超載保護:1000V 直流或交流峰值。 2-3-2.交流電壓(ACV) 1.將黑錶棒插入“COM"插孔,紅錶棒插入“ "插孔; "檔,初始為自動量程狀態,顯示“AUTO"符號 2.將旋鈕開關轉至“ 3.按動“SELECT"鍵選擇可手動選擇測量量程. 4.將測試錶棒接觸測試點,紅錶棒接的該點電壓顯示在螢幕上。 注意: 1. 測量電壓切勿超過 750V,如超過,有損壞儀錶電路的危險。 2. 當測量高電壓電路時,千萬注意避免人體觸及高壓電路,注意人身安全。 量程 準確度 分辨力 ±(0.
‧超載保護:1000V 直流或 750V 交流峰值; ‧頻率回應:(50~200)Hz ‧顯 示:平均值響應(以正弦波有效值校準)。 2-3-3.直流電流(DCA) A) 將黑錶棒插入“COM"插孔,紅錶棒插入“mA"插孔(最大為 400mA),或插入“10A"插孔中(最大為 10A) B) 將旋鈕開關轉至電流檔,儀錶初始為自動量程狀態,顯示“DC"符 號,然後將儀錶的錶棒串入被測電路中,被測電流值及紅色錶棒點的 電流極性將同時顯示在螢幕上。 注意: 如 LCD 顯示“OL",表明已超過量程範圍,需將量程開關轉至較高 一檔。 2. 最大輸入電流為 400mA 或者 10A(視紅錶棒插入位置而定) ,超過額 定的電流會將保險絲熔斷,甚至損壞儀錶。 量程 準確度 分辨力 400uA 0.1uA ±(1.0%+10d) 4000uA 1uA 40mA 10uA 400mA 100uA 4A ±(1.2%+10d) 10mA 10A ±(1.2%+10d) 10mA ‧最大測量壓降:滿量程 mA 為 0.4V,A 為 100mV; ‧最大輸入電流:10A(不超過 15 秒) ; ‧超載保護:0.
2. 最大輸入電流為 400mA 或者 10A(視紅錶棒插入位置而定) ,超過額 定的電流會將保險絲熔斷,甚至損壞儀錶。 量程 準確度 分辨力 400uA 0.1uA ±(1.5%+10d) 4000uA 1uA 40mA 10uA 400mA 100uA 4A ±(2.0%+15d) 10mA 10A ±(2.5%+15d) 10mA ‧最大測量壓降:滿量程 mA 為 0.4V,A 為 100mV; ‧最大輸入電流:10A(不超過 15 秒) 。 ‧超載保護:0.4A/250V 自恢復保險絲,10A/250V 保險管。 ‧頻率回應:10A 量程為(50~200)Hz。 2-3-5.電阻(Ω) 1. "插孔; 將黑錶棒插入“COM"插孔,紅錶棒插入“ A) 將旋鈕開關轉至“Ω"檔,兩錶棒跨接在被測電阻上。 B) 如果測量阻值小的電阻,應先將錶棒短路,測得引線電阻,然後在 實測中減去。 注意: 如 LCD 顯示“OL",表明已超過量程範圍,需將旋鈕開關轉至較高 一檔,當測量電阻超過 1MΩ 以上時,讀數幾秒之後才能穩定,這在 測量高阻時是正常的。 2. 當輸入端開路時,則顯示“OL"。 3.
‧開路電壓:小於 200mV; ‧超載保護:250V 直流或交流峰值; ‧注意事項:在使用 400Ω 量程時,應先將錶棒短路,測得引線電阻, 然後在實測中減去 警告:為了安全在此量程禁止輸入電壓值! 2-3-6.二極體及通斷測試 "插孔(注意紅錶棒極 A) 將黑錶棒插入“COM"插孔,紅錶棒入“ 性為“+"極) 。 B) 將旋鈕開關轉至““Ω"檔,按動“SELECT"鍵選擇二極體測量方式 C) 正向測量:將紅錶棒接到被測二極體的正極,黑錶棒接到被測二極體 的負極,顯示器即顯示二極體正向壓降的近似值。 D) 反向測量:將紅錶棒接到被測二極體的負極,黑錶棒接到被測二極體 的正極,顯示器顯示“OL"字樣。 E) 完整的二極體測試包括正反向測量,如果測試結果與上述不符,說明 二極體是壞的。 F) 按動“SELECT"鍵選擇通斷測量方式。 G) 將錶棒連接到待測線路的兩點,如果內置蜂鳴器發聲,則兩點之間電 阻值低於約 50Ω,則內置蜂鳴器發聲。 量程 顯示值 測試條件 二極體正向壓降 正向直流電流 0.5mA 反 向電壓約 1.5V 蜂鳴器發聲長響,測 開路電壓約 0.
注意: 1. 對被測電容應完全放電,以防止損壞儀錶; 2. 測量線上電容時,需將線路電源關斷,並將電容充分放電; 3. 200uF 量程輸入讀數穩定時間約 30 秒。 量程 準確度 分辨力 40nF ±(5.0%+30d) 10pF 400nF ±(3.5%+8d) 100pF 4μF ±(3.5%+8d) 1nF 20μF ±(3.5%+8d) 10nF 警告:為了安全在此量程禁止輸入電壓值! 2-3-8.頻率(Hz) 1. 測試電壓頻率時,將錶棒或遮罩電纜接入“COM"和“ 2. 將旋鈕開關轉至“ "輸入端; "檔,將錶棒或電纜跨接在信號源或被測負載 上,被測信號的頻率顯示在螢幕上。 注意: 輸入超過 10V 交流有效值時,可以讀數,但可能超差; 在雜訊環境下,測量小信號時最好使用屏壁電纜; 測量高電壓的頻率時,請選擇 ACV 檔,再按 Hz/%按鍵進入頻率測量 狀態 4. 禁止輸入超過 250V 直流或交流峰值的電壓,以免損壞儀錶。 程量 準確度 分辨力 ±(0.5%+10d) 400Hz 0.1Hz 4kHz 1Hz 20kHz 10Hz 0.1-99.9% 僅供參考 0.
注意: 1. 當輸入端開路時,則顯示常溫。 2. 請勿隨意更換測溫感測器,否則將不能保證測量準確度。 3. 嚴禁在溫度檔測量電壓。 準確度 分辨力 量程 1°C (-20~1000)°C <400°C±(1.0%+5d) ≥400°C±(1.5%+15d) 感測器:K 型熱電偶(鎳鉻—鎳硅)香蕉插頭。 警 告:為了安全在此量程禁止輸入電壓值! 2-3-10.晶體三極管 hFE 參數測試 1. 將功旋鈕開關轉至 HFE 檔位. 2. 將附件晶體管測試座插入 COM"和“mA"插孔中 3. 依據待測試的晶體管型號, 將各引腳插入對應的測試孔, 屏幕顯示相應 的測量數值。 量程 顯示值 測試條件 hFE NPN 或 PNP 0~1000 基礎電流約 10uA, Vce 約為 3V 警 告:為了安全在此量程禁止輸入電壓值! 七、儀錶保養 該儀錶是一台精密儀器,使用者不要隨意更改電路。 1. 請注意防水、防塵、防摔 2. 不宜在高溫高濕、易燃易爆和強磁場的環境下存放、使用儀錶 3. 請使用濕布和溫和的清潔劑清潔儀錶外表,不要使用研磨劑及酒精等 烈性溶劑 4.
1. 2. 3. 4. 注意: 不要將 1000V 直流或交流峰值電壓接入 不要在電流檔、電阻檔、二極體檔和蜂鳴器檔上,測量電壓值 在電池沒有裝好或後蓋沒有上緊時,請不要使用此錶 在更換電池或保險絲前,請將測試錶棒測試點移開,並關機。 八、故障排除 如果您的儀錶不能正常工作,下面的方法可以幫助您快速解決一般問題。 如果故障仍排除不了,請與維修中心或經銷商聯繫。 故障現象 檢 查 部 位 及 方 法 沒顯示 電源未接通,或更換電池 符號出現 顯示誤差大 更換電池 更換電池 本說明書如有改變,恕不另行通知 本說明書的內容被認為是正確的,若用戶發現有錯誤、遺漏等,請與 生產廠家聯繫 本公司不承擔由於用戶錯誤操作所引起的事故和危害 本說明書所講述的功能,不作為將產品用做特殊用途的理由。 ©2013 Prokit’s Industries Co., LTD.