MT-1820 3 5/6 Dual Display DMM w/USB Connector User’s Manual 1st Edition, 2013 ©2013 Copyright by Prokit’s Industries Co., Ltd.
Index General…………………………………………………………………. 2 Open Package Inspection…………………………………..……. 2 Safety Note………………………………………………………… 3 Safety Symbol Description…………………………………...4 Fr on t Pa nel Des c ri pti on …… …… … … …… …… … … ….. .5 Property………………………………………………………………….7 DC Voltage (DCV) …………………………………………………7 AC Voltage (ACV) …………………………………………………8 DC Current (DCA) …………………………………………………9 AC Current (ACA) …………………………………………………11 Resistance …………………………………………………12 Diode and Continuity Test ……………………………………...
General The instrument is a high performance, high accuracy, 3 5/6 dual display DMM w/usb connector multimeter, it uses the LCD with 24mm high figure to make the reading clear, the display simple and the operation convenient.
Safety Note The instrument meets the standard of IEC1010 (safety standard promulgated by the International Electrician Committee). Design and manufacture complied with the standard of Pollution Degree Warning To avoid endangering the users safety, should read the operation manual carefully before operation, and strictly obey the safety warning information and operation description when use the instrument. 1.
7. Do not store and use the instrument in high humility, high temperature, combustible, explosive and strong magnetic places. 8. It is possible to damage the instrument and endanger the safety of the users when measuring the voltage over the range limit. The allowed maximum voltage is printed on the front panel of the instrument, do not input the range limit specified to avoid the electric shock and instrument damage 9.
Front Panel Description 1. Instrument Model Number 2. LCD Display 3. RANGE: Auto Range/ Manual Range switch, the default is set as Auto Range mode when turning on, press the button and switch to Manual Range. At the Manual Range mode, press the button once, the range is switched to the higher one, press the button again to switch the range to the lowest one when measuring the highest range, the cycle is in proper order from low to high. Keep pressing the button for 2 sec, return to Auto Range mode.
gets into the MAX mode, the measuring maximum value will be held at this mode; Press the button again to get into the MIN mode, the measuring minimum value will be held at this mode. At the MAX/MIN mode, the maximum or minimum value will be held on LCD. Under this circumstance, there are no Bar Graph and Auto Power Off function, hold pressing MAX/MIN button for 2 seconds, exit the MAX/MIN mode. 7. SELECT: function switch, press the button to switch the measuring function. 8.
Property General Feature 1-1 Display: LCD 1-2 Max Display: 5999 (3 5/6) counts automatic polarity display and unit display. 1-3 Measuring method: dual-integral A/D converter 1-4 Sampling rage: 3 times/second 1-5 Over range indication: display “OL” 1-6 Low battery indication: “ ” appearance 1-7 Operation environment: 0~40˚C relative humidity <80% 1-8 Storage environment: -10~50˚C relative humidity <80% 1-9 Power: 2pcs 1.5V batteries (AAA battery) 1-10 Dimension: 185mm x 92mm x 48mm 1-11 Weight: Approx.
Note: a) Do not measure the voltage higher than DC 1000V or AC 750V. b) When measuring the high voltage, caution to avoid electric shock. Cut the connection between the test lead and tested circuit at once after measurement. C) At the Manual Range mode, if “OL” is displayed on LCD, it indicates the tested voltage value has exceeded the present range limit, please select the higher range to complete the measurement Range Accuracy Resolution 600mV 0.1mV 6V 1mV 60V ±(0.
the red test lead and the tested voltage value will be displayed on LCD simultaneously Note: a) Do not measure the voltage higher than DC 1000V or AC 750V. b) When measuring the high voltage, caution to avoid electric shock. Cut the connection between the test lead and tested circuit at once after measurement.
3. At the Manual Range mode,if “OL” is displayed on LCD, it indicates the test current value has exceeded the present range limit, please select the higher range to complete the measurement. Note: a) Do not measure the current higher than 10A at Range 10A and higher than 6000uA at uA Range and higher than 600mA at mA Range, otherwise the fuse will be burnt out or the instrument will be damaged.
Max input current: 10A (less than 15 seconds). Overload protection: 0.5A/250V fuse, 10A/250V fuse. AC Current (ACA) 1. Insert the black test lead to “COM” input terminal and the red test lead to “mA ”or “10A”input terminal,displayed as the picture. 2.
Range Accuracy Resolution 600uA 6000uA 60mA 600mA 6A 10A 0.1μA 1μA ±(1.5%+10d) 10μA 100μA 1mA 10mA ±(2.0%+15d) Max input current : 10A (less than 15 seconds). Frequency response: 40-100Hz Overload protection:0.5A/250V fuse, 10A/250V fuse. Duty Circle Indication: (0.1%-99.9%) Resistance (Ω) 1. Turn the function/range selection knob to Ω, insert the red and black test lead separately to VΩHz and COM input terminal. 2.
4. At the Manual Range mode,if “OL” is displayed on LCD, it indicates the tested resistance value has exceeded the present range limit, please select the higher range to complete the measurement. Note: a) When measuring the in-circuit resistance, Make sure all the power of the tested circuit has been turned off and all capacitors are fully discharged. b) It will cause the reading incorrect to input any voltage when measuring the resistance.
Open circuit voltage: 400mV. Over range protection: 250V AC/DC peak value. Diode and Continuity Test 1. Turn the function/range selection knob to Range, Insert the red and black test lead separately to VΩHz and COM input terminal. Press “SELECT” button to choose the Buzzer or Diode measurement function. 2. Connect the red test lead to the positive pole of the tested diode, BLACK test lead to the negative pole. 4. Read the present test result from LCD.
Range Display Diode Forward voltage drop Test Condition Forward AC Current:1.0mA, Counter Voltage: 3.0V Buzzer sound at less than Open Voltage: 0.5V 50Ω Overload protection: 250V DC/AC peak value. Warning: Do not input the voltage at the range for safety. Capacitance (C) 1.Turn the function/range selection knob to Capacitance Range. Insert the red and black test lead separately to VΩHz and COM input terminal. 2.
once after measurement. Range Accuracy 40nF ±(5.0%+30d) 400nF 4μF 10pF 100pF ±(3.5%+8d) 40μF 200μF Resolution 1nF 10nF ±(5.0%+10d) 100nF Overload protection: 250V DC/AC peak value WARNING: Do not input any voltage at this range for safety! Frequency (Hz) 1. Turn the function/range selection knob to Hz Range. Insert the red and black test lead separately to VΩHz and COM input terminal. 2. Connect the test lead to the tested signal source in parallel, Read the present test result from LCD.
3. When measuring Frequency, press “Hz/DUTY” button once to get into the mode of DUTY measurement, and press “Hz/DUTY” button again to return to the mode of Frequency measurement. 4. When measuring AC current or voltage, press “Hz/DUTY” button to get into the mode of Frequency measurement, and press “Hz/DUTY” button again to get into the mode of Duty Circle measurement, and press the button third to return to the mode of AC current or voltage measurement. Note: a) Do not input the signal more than 60V.
Transistor Test 1. Turn the function/range selection knob to “hFE” Range, displayed as the right picture. 2. Insert the accessory of Transistor test into the “mA ” and “COM” input terminal, displayed as the picture.According to the model of the tested transistor, insert the accessory accordingly into the terminal of “EBC” of “NPN” or “PNP”. 4. Read the present test result from LCD.
positive end into the “mA” input terminal. 3. Connect the sensor of the temperature cable to the surface or inside of the tested object, read the present test result from LCD. 4. Press “SELECT” button to switch Fahrenheit or Centigrade temperature measurement mode. Note: a) Without the signal input, LCD automatically displays the present temperature. b) Do not replace the temperature sensor at will. Otherwise the accuracy can’t be assured.
Other Functions and Instrument Maintenance • Other Functions 1. Data Hold Press “HOLD” button, the present value is held on LCD; Press the button again, exit the hold mode and get in the normal measuring status. 2. Auto Power Off Stop working for 15 minutes, the instrument will auto power off, and get into the sleeping mode. The buzzer inside will sound 5 times within one minute before powering off, there is a long sound one minute later and the instrument gets into the sleeping mode.
Instrument Maintenance This is a highly precise instrument, do not try to modify the inner circuit at will. 1. Keep the instrument dry, and keep it away form dust and shock. 2. Do not store and use the instrument in high humility, high temperature, combustible, explosive and strong magnetic places. 3. Clean the surface of the instrument with the damp cloth and gentle detergent, do not use the strong solvent like the abrasive cleaner and alcohol, etc. 4.
tightening the bottom case 4. Please remove the test lead from the tested point and turn the power off before replacing the battery or fuse. FAULT ELIMINATION If the instrument does not work properly, the below methods can help you to solve the problems quickly.
MT-1820 3 5/6 USB 連線型雙顯自動錶 目 錄 概述 ……………………………………………………………..24 開箱檢查 …………………………………………………………… 24 安全注意事項…………………………………………………24 安全符號說明 …………………………………………………26 儀錶面板及按鍵功能說明 ………………………………………… 26 其他功能…………………………………………………28 特性…………………………………………………….. 28 直流電壓(DCV) ………………………………………………… 29 交流電壓(ACV) ………………………………………………… 30 直流電流(DCA) ………………………………………………… 31 交流電流(ACA) …………………………………………………32 電阻…………………………………………………………………. 34 二極體及通斷測試 ………………………………………………… 35 電容(C) ……………………………………………………………..36 頻率(Hz) ……………………………………………………………..
一.概述 MT-1820 一種性能穩定、高可靠性 3 5/6 USB 連線型雙顯自動錶,儀 錶採用 24mm 字高 LCD 顯示器,讀數清晰,顯示直觀,操作方便, 可用來測量直流電壓、交流電壓、直流電流、交流電流、電阻、電容、 頻率、占空比、電晶體、二極體及通斷測試;同時還設計有 61 段類 比棒條顯示、單位符號顯示、資料保持、相對值測量(REL),帶 USB 電腦介面,自動/手動量程轉換、自動斷電及報警功能。整機採用直接 驅動 LCD 的 4 位元微處理器和雙積分 A/D 轉換積體電路,提供高分 辨力、高精度的數位顯示驅動,該表功能齊全,測量準確度高,使用 方便,是實驗室、工廠、無線電愛好者及家庭的理想工具。 二、開箱檢查 打開包裝箱取出儀錶,仔細檢查以下附件是否缺少或損壞,如有缺少 或損壞請立即與經銷商聯繫。 •數字多用表 一台 •使用說明書 一本 •表棒 一副 •溫度測試線(K 型熱電偶) 一條 •USB 連接線 一條 •軟體光碟 一張 •電晶體測試座 一個 三、安全注意事項 MT-1820 符合 IEC1010 條款(國際電工委員會頒佈的安全標準) 。使 用污染等級 2
為避免危險及使用者的安全,在使用儀錶之前請仔細閱讀本使用手 冊,並嚴格遵守安全警告資訊和操作說明來使用本儀錶。 在測量 30V 以上電壓,測量 10mA 以上電流,測量帶電感負載的交 流電力線,測量電力波動期間的交流電力線時,謹防電擊。 2.測量前,檢查功能/量程選擇開關是否置於正確的檔位,要檢查表棒 是否可靠接觸,是否正確連接、是否絕緣良好等,以避免電擊。 3.儀錶只有和所配備的表棒一起使用才符合安全標準要求。如表棒線 破損時,必需更換上同樣型號或者相同電氣規格的表棒線。 4.不要使用其他未經確認或未認可的保險管來更換儀錶內部的保險 管,只能換上同樣型號或相同規格的保險管,更換前表棒必需離開被 測量點,確保輸入端無任何信號。 5.不要使用其他未經確認或未認可的電池來更換儀錶內的電池,只能 換上同型號或相同電氣規格的電池,更換前表棒必需離開被測量點, 確保輸入端無任何信號。 6.在進行電氣測量時,身體切勿直接接觸大地,不要接觸可能存在地 電勢裸露的金屬端子、輸出口、引線夾等。通常使用的乾燥的衣服、 膠鞋、膠墊以及其他絕緣材料,保持你的身體與大地絕緣。 7.
12.當 LCD 顯示“ ”時,請及時更換電池以確保測量精度。 13.不允許表棒插在電流端子去測量電壓! 14.請不要隨意改變儀錶線路,以免損壞儀錶和危及安全。 四、安全符號說明 警告! 直流 高壓!危險! 交流 大地 交直流 雙重絕緣 符合歐洲工會指令 電池欠壓 保險絲 五、儀錶面板及按鍵功能說明 1. 儀錶型號 2. 61 段類比棒條 LCD 屏顯示。 3. RANGE:自動/手動量程切換,開機時預設為自動量程,按一下 切換為手動量程,在手動量程模式下,每按一下往上跳一檔,到 最高檔時再按此鍵又跳到最低檔,依次循環。如按此鍵超過 2 秒則切換回到自動量程。頻率、fHE 及電容檔沒有手動量程功能。 4.
5. Hz/DUTY: 頻率/占空比選擇鍵, 在頻率檔按此鍵可以切換頻率或 占空比功能;在交流電壓或交流電 流檔位時,按此鍵可以在電壓或電 流/頻率/占空比之間作切換。 6. MAX/MIN : 最大值/最小值顯示, 按下此功能鍵,進入 MAX 模式保 持測量的最大值:再按一次進入 MIN 模式保持測量的最小值,在此 工作環境下無類比棒條顯示和自 動關機功能,按 MAX/MIN 鍵 2 秒 後,退出 MAX 或 MIN 測量測試。 7. SELECT : 按鍵開關,用於選擇各 種測量功能切換。 8. RS232/REL:相對值測量,按一 下此鍵為相對值測量,再按一次取消相對值測量,以此循環,如 按此鍵超過 2 秒則切換到資料傳輸 PC 模式,LCD 屏上有 RS232 顯示,此時電錶資料可輸出至 PC,如再按此鍵超過 2 秒則 RS232 在 LCD 屏上消失,此時關閉 RS232 資料輸出。 9. 功能/量程選擇開關 : 用於選擇各種測量功能和量程。 10. 10A 電流輸入插孔:測量交直流電流 10A 檔的正輸入端,插入 紅表棒。 11.
六、其他功能 1.自動斷電 當儀錶停止使用 15 分鐘後,儀錶便自動斷電(關機) ,然後進入睡眠 (關機)狀態,斷電前一分鐘內置蜂鳴器會發出 5 聲提示,一分鐘後 長響一聲進入休眠(關機)狀態,若要重新啟動電源(開機),請按 任意鍵或撥動功能/量程選擇開關,均可重新開機。 2. 按“RS232/REL”鍵,為相對值測量,常按“RS232/REL”鍵,大於兩 秒,顯示幕上有 RS232 顯示,此時儀錶可與電腦連接,利用所附的 USB 連接線連接儀錶及電腦,可將所測量的資料進行記錄、分析、 處理和列印。詳細請參考軟體中的說明。 3. 背光照明功能: 若需要在昏暗環境下進行測量,請按下”HOLD”鍵 2 秒, 背光照明開啟,持續大約 10 秒後關閉.
技術特性 •準確度:±(讀數的% + d 位數),保證準確度環境溫度 23±5℃,相 對濕度<75%, •校準保證期從出廠日起為一年。 八、直流電壓(DCV) 1. 將功能/量程開關旋至交/直流電壓,紅黑表棒分別插入“VΩHz”和 “COM”孔中。 2. 儀錶的起始狀態為自動量程直流電壓模式,LCD 顯示“AUTO”符 號,按” RANGE“鍵,可轉換為手動量程模式。 3. 將表棒接觸測試點並聯到被測電路上,紅色表棒線的極性和被測 電壓值將同時顯示在 LCD 營幕上。 注 意: a) 不能測量高於 DC1000V 或 AC750V 的電壓。 b) 在測量高壓時,特別注意避免觸電。在測試完後立即斷開表棒與 被測電路。 c) 手動量程模式下,如營幕上顯示“OL”,則被測電路已經超過所選 量程,請選擇更高的量程檔位來完成此次測量。 量 程 準確度 600mV 分辨率 0.1mV 6V 1mV ±(0.5%+4d) 60V 10mV 600V 100mV 1000V ±(1.
輸入阻抗:600mV 量程>60MΩ,其餘為 10MΩ。 超載保護:1000V 直流或 750V 交流峰值。 九、交流電壓(ACV) 1.將功能/量程開關旋至交/直流電壓,按”SELECT”選擇交流電壓,將 紅黑表棒分別插入“VΩHz”和“COM”。 2.儀錶的起始狀態為直流電壓模式,按”SELECT”鍵可切換為交流電 壓模式,LCD 初始為自動量程顯示“AUTO”符號,按“RANGE“鍵,可 轉換為手動量程模式。 按下“Hz/DUTY”鍵可測量頻率/占空比測量,但此時頻率回應較低,適 用於磁場干擾環境下的高壓低頻測量,如 220V/50Hz-400Hz, 380V/50Hz~400Hz。 3.
量 程 準確度 分辨率 6V 60V 1mV ±(0.8%+10d) 600V 750V 10mV 100mV ±(1.0%+10d) 1V 輸入阻抗: 10MΩ。超載保護:1000V 直流或 750V 交流峰值。 頻率回應: 40~400Hz,顯示: 平均值回應(以正弦波有效值校準)。 占空比顯示:(0.1%-99.9%)。 十、直流電流(DCA) 1.將黑表棒插入“COM”孔中,紅表棒插入“mA ”或“10A”孔中,如下圖 所示。 2.將功能/量程開關轉至所需電流檔,LCD 顯示“AUTO”符號,初始值 均設定為直流測量方式, 按“RANGE“鍵,可轉換為手動量程模式。 3.將表棒接觸測試點串聯到被測電路上,被測電流值及紅表棒點的電 流極性將同時顯示在螢幕上。 4.
量 程 準確度 分辨率 60OuA 0.1μA 60O0uA 1μA 60mA ±(1.0%+10d) 600mA 6A 10A 10μA 100μA ±(1.2%+10d) 1mA 10mA 最大輸入電流:10A(不超過 15 秒) 超載保護:0.5A/250V 保險絲;10A/250V 保險絲。 十一、交流電流(ACA) 1.將黑表棒插入“COM”孔中,紅表筆插入“mA ”或“10A”孔中,如右圖 所示。 2.將功能/量程開關轉至所需電流檔,按“SELECT”鍵,選擇交流測量 方式,將表棒接觸測試點串聯到被測電路上,被測電流值及紅表筆點 的電流極性將同時顯示在螢幕上 。 3.
注 意: a) 在 10A 檔不能測量大於 10A 的電流,“uA”檔不能測量大於 6000uA 的電流,“mA”檔不能測量大於 600mA 的電流。 b)當表棒插在電流輸入端時,嚴禁將表棒並聯在任何電路上,否則有 可能損壞儀錶和危及使用者安全,在完成所有得測量後,要立即斷開 表棒與被測電路。 量 程 準確度 分辨率 600uA 6000uA 60mA 0.1μA 1μA ±(1.5%+10d 10μA 600mA 6A 10A 100μA 1mA ±(2.0%+15d) 10mA 最大輸入電流:10A(不超過 15 秒) 。 超載保護:0.5A/250V 保險絲;10A/250V 保險絲。 頻率回應: 40~100Hz。 占空比顯示:(0.1%~99.
十二、電阻(Ω) 1.將功能/量程開關旋至“Ω”檔,將紅黑表棒分別放入“VΩHz”和 “COM”。 2.將表棒的測試端並聯到被測量電阻上,被測電阻值會顯示在營幕 上。 3.開機預設為自動量程,LCD 顯示“AUTO”符號,按“RANGE”鍵選擇 手動量程模式。 4.
量 程 600Ω 準確度 分辨率 ±(0.8%+5d) 0.1Ω 6kΩ 60kΩ 600kΩ 1Ω 10Ω ±(0.8%+4d) 100Ω 6MΩ 60MΩ 1kΩ ±(1.2%+10d) 10kΩ 開路電壓:400mV。 超載保護:250V 直流或交流峰值。 十三、二極體及通斷測試 將功能/量程選擇開關旋到“ ”檔。 2. 將紅、黑色表棒分別插入“VΩHz”和“COM”輸入端,再按“SELECT” 選擇二極體或通斷測量。 3.將紅色表棒接到二極體的正極,黑色表筆接到二極體的負極。 4.
量 程 顯 示 值 測 試 條 件 正 向 直 流 電 流 約 1.0mA , 反 向 電 壓 約 二極體正向壓降 3.0V 峰鳴器發聲長響,測試二點 電阻約 50Ω 開路電壓約 0.5V 超載保護:250V 直流或交流峰值。 警 告:為了安全在此量程禁止輸入電壓值! 十四、電容(C) 1.將功能開/量程開關旋到電容檔。將紅黑表棒分別插入“VΩHz”和 “COM”輸入端。 2.將表棒的測試端並聯到被測電容上,被測電容值將顯示在營幕上。 3.
注 意: a) 在測電容時,必須先將被測線路內所有的電源斷開,並將所有的 電容充分放電。注意:該檔無類比棒條顯示) b) 當測量大電容時,需要較長的時間,在 100uF 時約 15 秒。 C) 完成所有的測量後,要立即斷開表棒與被測電路的連接。 量 程 40nF 準確度 分辨率 ±(5.0%+30d) 10pF 400nF 4μF 100pF ±(3.5%+8d) 1nF 40μF 200μF 10nF ±(5.
十五、頻率(Hz) 1. 將功能/量程開關旋到“Hz”檔。 將紅黑表棒分別插入“VΩHz”和 “COM”輸入端。 2. 將表棒的測試端並聯到待測信 號源上,從營幕上讀出結果(注 意:該檔無類比棒條顯示) 。 3. 在進行頻率測量時,按一次 “Hz/DUTY”鍵進入占空比測量狀 態,再按一次“Hz/DUTY”鍵進入頻 率測量狀態。 4. 當功能/量程開關旋到交流電流/交流電壓檔測量時,按“Hz/DUTY” 鍵進入頻率測量狀態,再按一次“Hz/DUTY”鍵進入占空比測量狀態, 第三次按“Hz/DUTY”鍵返回到原測量狀態。 注 意: a) 不要輸入高於 60V 的信號,否則可能損壞儀錶並危及人身 安全。 b) 在完成所有的測量後,要立即斷開表棒與被測電路。 量 程 準確度 分辨率 100Hz 0.01Hz 1000Hz 1Hz 10kHz 10Hz 100kHz ±(0.
輸入靈敏度:1.0V。 超載保護:250V 直流或交流峰值。 占空比顯示:(0.1%~99.9%)。 十六、電晶體測量 1. 將功能/量程開關旋到“hFE”檔如右圖。 2. 將電晶體附件如圖插入“mA ”與“COM”孔中。 3. 根據被測電晶體的型號,插入電晶體附件所對應的:“NPN”或 “PNP”的“E B C ”孔中。 4.
量 程 hFE(NPN 或 PNP) 顯 示 範 圍 0-1000 測 試 條 件 基極電流約 1mA VCE 約 2.1V 十七、溫度測量 1.將功能/量程選擇開關旋到溫度檔。 2.將溫度探頭的輸入端插入“COM”,正端插入“mA”中,將溫度探頭的 感應端置於被測物件的表面上,如下圖的水中。 3.
量程 準確度 分辨率 -20℃-400℃ ±(1.0%+50d) 0.1℃ 400℃-1000℃ ±(1.5%+15d) 1℃ 0℉-750℉ ±(1.0%+50d) 0.1℉ 750℉-1832℉ ±(1.5%+5d) 1℉ K 型熱電偶(鎳鉻-鎳硅)香蕉探頭。 警 告:為了安全在此量程禁止輸入電壓值! 十八、通訊連接 1.按包裝所提供的光碟對應儀錶的型號,選擇 setup 的文件安裝。 2.將儀錶通過 USB 連接線與電腦連接好。 3.長按“RS232/REL”兩秒以上,螢幕上出現 RS232 顯示。 4.測量的資料傳輸到電腦,此時可以對資料進行記錄、分析、處理和 列印等。 詳細請參考軟體中的說明。 十九、儀錶保養 該儀錶是一台精密儀器,使用者不要隨意更改電路。 1.請遠離有水氣及灰塵之處。請勿重摔。 2.請勿在高溫高濕、易燃易爆和強磁場的環境下存放、使用儀錶。 3.請使用濕布和溫和的清潔劑清潔儀錶外表,勿使用研磨劑及酒精等 烈性溶劑。 4.如果長時間不使用,應取出電池,防止電池漏液腐蝕儀錶。 5.
步驟如下: a.取出後蓋上固定電池的螺絲,取下電池蓋。 b.取下 2 個 1.5V 電池,換上兩個新的電池,雖然任何標準 1.5V 電池 都可使用,但為加長使用時間,最好使用鹼性電池; c.裝上電池蓋,鎖緊螺絲。 d.保險絲更換:步驟同上。更換保險絲時,請使用規格、型號相同的 保險絲。 二十、故障排除 如果您的儀錶不能正常工作,下面的方法可以幫助您快速解決一般問 題。如故障仍無法排除,請與維修中心或經銷商聯繫。 故障現象 沒顯示 符號出現 檢查部位及方法 電源未接通 保持開關 換電池 換電池 電流沒輸入 換保險絲 顯示誤差大 換電池 注 意: 1.不要將高於 1000V 直流或交流峰值電壓接入。 2.不要在電流檔、電阻檔、二極體檔和蜂鳴器檔測量電壓值。 3.在電池沒有裝好或後蓋沒有上緊時,請不要使用此表。 4.