MT-5110 3 1/2 Capacitance Meter User’s Manual st , 1 Edition 2012 ©2012 Copyright by Prokit’s Industries Co., Ltd.
1. FEATURES ◇ ◇ ◇ ◇ ◇ ◇ ◇ ◇ ◇ Easy and correct readout. High measuring accuracy. Measurements are possible even under a strong magnetic field. LSI-circuit provides high reliability and durability. Input overload protection is provided. LCD display for clear readout even in bright ambient light conditions. Light-weight and compact construction for easy operation. Low battery condition is indicated on the LCD display. Low power consumption. 2. SPECIFICATIONS 2-1.
2-2. ELECTRICAL SPECIFICATION Accuracy is ±(percentage of reading + number of digit) at 23±5 ℃,<80%RH. Range Accuracy Resolution Test Frequency Max indication value 200pF ±(0.5%+7) 0.1pF 800Hz 199.9pF 1pF 10pF 100pF 1000pF 0.01uF 0.1uF 1uF 10uF 800Hz 800Hz 800Hz 800Hz 80Hz 8Hz 8Hz 8Hz 1.999nF 19.99nF 199.9nF 1.999uF 19.99uF 199.9uF 1999uF 1999(×10)uF 2nF 20nF 200nF 2uF 20uF 200uF 2000uF 20000uF ±(0.5%+5) ±(2%+5) ±(3%+10) -12 -9 -6 pF= Pico Farad(10 F ),nF= nan Farad(10 F).
4. CONSIDERATION OF MEASUREMENT (1) This C METER is intended for measuring the capacitance value of a capacitor. It is not intended for determining the “Q” factor for above reactive components. Misleading readings may be obtained if the measurement of capacitance of a resistor is attempted. (2) When measuring components within circuit that circuit must be switched off and de-energized before connecting the test leads.
NOTE: (a) If the capacitance value is unmarked, start with the 200pF range and keep increasing until the over-range indication goes off and a reading is obtained. (b) A shorted capacitor will read over-range on all ranges. A capacitance with low voltage leakage will read over range, or a much higher value than normal. An open capacitor will read zero on all ranges (possibly a few pF on 200pF range, due to stray capacitance of the instrument).
6. MAINTENANCE 1) 9-Volt battery replacement a. Ensure the instrument is not connected to any external circuit. Set the selector switch to OFF position and remove the test leads from terminals. b. Remove the screw on the bottom case and lift the bottom case. c. Remove the spent battery and replace it with a battery of the same type. 2) Fuse replacement a. Ensure the instrument is not connected to any external circuit. Set the selector switch to OFF position and remove the test leads from terminals. b.
MT-5110 3 1/2 數位電容錶 1. 特色 ◇ 測量準確, 讀數方便 ◇ 測量精度高 ◇ 在幹擾的磁場下依然可以測量 ◇ LSI 電路提供高可靠性和耐久性 ◇ 提供輸入過載保護 ◇ 即使在明亮室內光線狀況下, LCD 清楚顯示測量結果 ◇ 結構設計容易操作使用 ◇ 若工作電池電壓低, LCD 上會顯示低電壓符號提醒 ◇ 整機低功耗 2. 說明 2-1.一般說明 最大讀值 1999 顯示︰LCD(液晶顯示螢幕) 測量︰C(僅電容量) 範圍︰9 個檔位, 全範圍值從 0.1pF to 20000uF 歸零調整︰手動(範圍︰大約 ±20pF) 校驗調整︰內含 2 個調整裝置 輸入超過︰螢幕顯示 “1” 背光功能︰ 8 秒內 採樣時間︰0 ~5second 操作溫度︰0 ℃~40 ℃.最大濕度:80%.R.H.
2-2. 電氣規格 精度為 ±(讀數的百分比+個位數)在 23 ± 5℃,<80%相對濕度。 範圍 精準度 解析度 測試頻率 最大顯示值 ±(0.5%+7) 200pF 0.1pF 800Hz 199.9pF 2nF 1pF 800Hz 1.999nF 20nF 10pF 800Hz 19.99nF 200nF 100pF 800Hz 199.9nF ±(0.5%+5) 2uF 1000pF 800Hz 1.999uF 20uF 0.01uF 80Hz 19.99uF 200uF 0.1uF 8Hz 199.9uF 2000uF ±(2%+5) 1uF 8Hz 1999uF 20000uF ±(3%+10) 10uF 8Hz 1999(×10)uF -12 -9 -6 pF= Pico Farad(10 F ),nF= nan Farad(10 F). uF= micro Farad(10 F) Excitative voltage: Max.2.8Vrms 功率超載:由 0.1A/250V 保險絲保護。 3. 操作面板 1.產品型號. 2.LCD 螢幕︰顯示測試值和單位 3.
(3) 注意檢查(黑色和紅色)測試線, 若損壞請更換. (4) 在佈滿灰塵的環境裡使用儀器應該定期進行清潔。 (5) 不要長期把儀器暴露在太陽下。 (6) 取出電池和打開保險絲蓋子之前,確認儀器任何連接或電源開關處於 關閉位置。 (7) 所有的測量,應該連接到黑色測試線 “ - ”和紅色測試線 “+”。 5. 電容(C)測量步驟 (1) 將功能旋鈕轉到待測試電容的預期最大值選擇的範圍 (2) 電源會同步開啟。 (3) 檢查“0”位:如果測試的範圍是 200pF,2nF,20nF,應測試前檢查 儀表顯示為“0”, 若有偏離請調整”ADJ〞調零旋鈕. (4) 當測試有極性電容器時,檢查其極性, 以對應測試插座 “+”端,”-” 端連接.
(e) 電容器,特別是電解電容,往往有廣泛的公差。不要感到驚訝,如 果測量值大於標在電容器上的值,除非它是一個緊公差,不管如何 測量,測量值很少大幅低於額定值。 (f) 如果更改測量檔位,測量顯示值將不同。 電容器漏電壓也將隨檔位 不同而變化, 在低檔測量位置.漏電電阻會變小。 6. 維修 1) 9 伏特的電池更換 a. 確認儀器不連接任何外部電路。確定儀器功能旋鈕轉向為 OFF 位置並 且除去測試線。 b. 取下底蓋的螺絲與蓋子。 c. 取出電池並更換一個同類型的電池。 2) 保險絲更換 a. 確認儀器不連接任何外部電路。確定儀器功能旋鈕轉向為 OFF 位置並 且除去測試線。 b. 取下底蓋的螺絲與蓋子。 c.
©2012 Prokit’s Industries Co., LTD.