Service manual

METTLER TOLEDO 
DIVA Service Manual  45 
6.1.4.3  Status Byte Format 
When the scale responds with STX?status byte 
C
R
 , this indicates a status byte. The status byte 
contains scale status information, which can include net/gross mode, zero status, etc. The actual 
status byte sent will be an ASCII character which must be converted to binary to decode the bits. 
The scale status bit definitions are shown in the following table. 
Scale Status Byte Bit Definitions 
Bit Number  Bit Description 
7 
7 Parity Bit (even). 
6 
Always 1 
5 
1 = Net Weight,   
0 = Gross Weight 
4 
1 = Center of zero.   
0 = Not at center of zero. 
3 
1 = Outside zero capture range.   
0 = Within range. 
2 
1 = Under zero.   
0 = Within weighing range. 
1 
1 = Over capacity.   
0 = Within weighing range. 
0 
1 = Scale in motion.   
0 = Stable weight data. 