Specifications
5-2 Magellan
TM
 8500Xt Scanner
Description of Calibration Sequence
The Calibration Sequence sets the scale to an accurate reference point for 
weighing. This process involves the use of a Field Standard Weight Set 
(31.5-pounds) for calibration in pounds, (18.5-kilograms) for Metric. 
Once calibration has been successfully completed, the scanner/scale uses 
the certified weight as a reference for subsequent weighing activities.
These verification procedures follow the U.S. 
National Institute of Stan-
dards and Technology 44 Handbook guidelines for 
bench/counter scale 
installations.
If any of these tests fail to meet the required weight indications, you must 
calibrate th
e scanner/scale. Refer to the calibration procedures in this sec-
tion for the proper procedures.
You may be required by state and/or local
 regulations to have procedures 
other than these performed by a certified technician or verification offi-
cial.
Access to the calibration switch should be restricted with a paper or a wire 
and lead seal 
after the calibration has been performed if required by your 
local regulatory agency.
NOTE
The Calibration Sequence must be perf
ormed without removing the scanner/
scale from its installed position.
The following tools and supplies will be required to perform the calibra-
tion process:
• 31.5 pound Field Standard Weight Set
1
. 
(English calibration only).
• 18.5 kilogram Field Standard Weight Set
2
. 
(Metric calibration only).
• Lead/Wire or Paper Seal (as required by law).
1. NOTE: Throughout the calibration procedures, specified weights may be achieved by using a com-
bination of weights from this set. eg.
 20 lbs may be made up of one 10.00 lb. weight and two 5.00 
lb. weights.
2. NOTE: Throughout the calibration procedures, specific weights may be achieved by using a com-
bination of weights from this set. eg.
 10 kgs. may be made up of one 5.00 kg. and five 1.00 kg. 
weights.