12  Technical specifications
Classification according to Directive 93/42/EEC: 
Class I (one)
Safety Standards:  The product has been tested and found to comply to IEC 60601-1 
(second edition) and to IEC 60601-1-1.
  The product has been tested and found to comply to IEC 60601-1 
(third edition) and to IEC 60601-1-2 (third edition – for electro-
magnetic compatibility – EMC).
  The hand piece is intended for use within the patient environment 
whereas the charger unit and the power supply adapter are not 
intended for use within the patient environment.
Note: The patient environment is defined as an area 1.5 m from the 
patient (as per IEC 60601-1, third edition and IEC 60601-1-1).
Classification according to IEC 60601-1:  Class II, Type B, Applied Part: Light guide with protective cone.
  IP 20 (Charging unit).
  IP 20 (Translux Wave)
Charging unit:  Translux Wave Charging unit
Power supply for the charging unit:  Switching Power Supply (Direct Plug-In) manufactured by: 
Phihong Technology CO LTD – 
Model: PSACO5R-050. 
Input: 100 – 240 V   – 300 mA – 50 / 60 Hz –.
Output: 5 V   – 1 A Max –   .
  Class II equipment
Power supply for the hand piece:  Rechargeable Lithium-ion battery.
Manufacturer: Panasonic.
Model: NCR-18500.
Nominal Voltage: 3.6 V. 
Nominal capacity (Typical): 2000 mAh 
Hand piece:  Model: Translux Wave. 
Operation: Intermittent service: 120“ ON – 40“ OFF – max 2 times 
running
Light source:  High-luminosity LED.
  Class 2M LED Product (IEC 60825-1).
Wavelength range: 440 – 480 nm. 
Dominant wavelength: Min. 440 nm – Max. 465 nm
Light guide:  Lead-free multi core rod, 8 mm diameter.
Sterilisable in pre-vacuum steam autoclave: 
4 minutes (minimum) at minimum temperature of 132°C + 3°C 
(270°F + 5°F). Max. 500 cycles
Settable exposure cycles:
– Normal exposure
  10 seconds of exposure time. 
Acoustical signals:
  1 beep at the start of exposure.
1 beep at the end of exposure
– “Soft-Start” exposure  20 seconds of exposure time. (The “Soft-Start” means increase 
of light intensity from 50% up to 100% within 2 seconds.)
Acoustical signals:
  1 beep at the start of exposure. 
1 beep after 1 second for indication of “Soft-Start”-mode activation.
1 beep after 10 seconds of exposure.
1 beep at the end of exposure
Charging time of empty battery:  Approx. 3 hours
Environmental conditions of unit: 
Operation Transportation and 
storage
Temperature 10°C – 35°C 
(50°F – 95°F)
–20°C – 40°C 
(–4°F – 104°F)
Relative humidity 45% – 85% 45% – 85%
Atmospheric pressure 800 – 1060 hPa 500 – 1060 hPa