SECTION 2 SAFETY
INSTRUCTIONS
2.1 INSTALLATION SAFETY
Before starting or installing the TOC-PG3 TOC Gas Generator,
be sure that all power to the unit is off, valves are shut, and the
air circuit is at atmospheric pressure. DO NOT remove, repair,
or replace any component, control filter, or part, while the unit is
energized or the air circuit is under pressure. Unplug unit and
de-pressurize the unit before starting installation or
maintenance procedures.
The TOC-PG3 is rated NEMA 1.
2.2 OPERATION SAFETY
DO NOT OPERATE THE TOC GAS GENERATOR AT
COMPRESSED AIR PRESSURES ABOVE 120 PSIG. UNIT
FAILURE, INJURY AND EQUIPMENT DAMAGE COULD
RESULT
CAUTION:
EXCEPT as otherwise specified by the manufacturer, this
product is specifically designed for compressed air service and
use with any other gas or liquid is a misapplication.
Manufacturer’s warranties are void in the event of a
misapplication and manufacturer assumes NO
RESPONSIBILITY for any resulting loss.
Before using equipment with fluids or gases other than air,
consult Puregas for written approval.
SECTION 3 SPECIFICATIONS
3.1 FLOWS, CONNECTIONS, DIMENSIONS, AND WEIGHT
MODEL NO. TOC-PG3
Max Flow Capacities at 100 PSIG (L/M)
Inlet Flow 13.2
Purge Flow 5.5
Outlet Flow 7.7
Connections (NPT)
Inlet 1/4”
Outlet 1/4”
Dimensions (Inches)
Height 10.5
Width 12.5
Depth 4.5
Weight (Pounds)
6.5
Note: Inlet flows and outlet flows shown above are maximum capacities and should not be exceeded for best performance.
No outlet flow metering valve or flow meter is installed on the adsorber/dryer and must be provided by the user. Purge flow is
metered by an integral fixed purge orifice. Capacities are based on inlet conditions of 70°F, 100%RH, normal ambient CO
2
levels of approx 350-375 ppm. Outlet concentration of CO
2
will be less than 1 ppm and outlet dew point better than -100°F. If
your flow, temperature or pressure are different from above, consult factory for performance.
OUTLET FLOW
CAPACITIES AT
VARIOUS INLET
PRESSURES
PSIG lit/min
120 9.1
110 8.4
100 7.7
90 6.2
80 4.6
70 3.3
60 2.1
50 1.2