SECTION 4 INSTALLATION
4.1 PARTS INCLUDED
(1) TOC-PG3 TOC Gas Generator Assembly
(1) Operations & Maintenance Manual
4.2 LOCATION
DO NOT INSTALL THE UNIT IN AN ENVIRONMENT OF
CORROSIVE CHEMICALS, EXPLOSIVE GASES, OR AREAS
OF HIGH AMBIENT TEMPERATURE CONDITIONS.
INSTALL THE UNIT INDOORS.
4.3 ELECTRICAL CONNECTIONS
Before plugging in unit, check the unit nameplate for electrical
characteristics. Standard electrical characteristics are 115 volt,
1 phase, 50/60 Hz.
IMPORTANT!: No overload protection is provided in the
adsorber/dryer and unit should be plugged/wired into a
protected circuit.
4.4 PIPING AND CONNECTIONS
Inlet and outlet tubing should be appropriate for pressure and
temperature of operation. See specification chart for
inlet/outlet NPT thread size. Teflon tape should be used in the
makeup of joints to ensure a good, airtight fit of piping
components. Check all connections for leakage using soap
solution prior to putting unit into service.
4.5 FILTERS
A coalescing filter is provided at the inlet to the adsorber/dryer
to remove entrained particulates, liquid moisture and oil which
can cause damage to the desiccant beds. A particulate
afterfilter is provided after the adsorber/dryer to remove any
adsorbent media dust that may migrate from the desiccant
beds.
INLET COALESCING
FILTER WITH AUTO
DRAIN
OUTLET
PARTICULATE FILTER
-REGULATOR WITH
GAUGE
WALL MOUNTING
HOLES (x4)
COVER MOUNTING
HOLES (x6)