Use and Care Guide

9
QUARTZ SLEEVE:
Installation of the Quartz Sleeve:
Always handle quartz sleeves carefully to prevent breaking or chipping. The quartz sleeves are to be clean and
free of fingerprints before installing.
Remove the compression nut. Install the quartz sleeve (domed end first) into the stainless steel housing, through
the threaded nipple. The quartz sleeve will go to the bottom of the chamber and through a quartz receiver locat-
ed within the bottom of the UV chamber. This will leave approximately 1/2" of the quartz sleeve exposed above
the SS nipple. (If you have more then this amount or it appears to have a slight tilt, the sleeve is NOT seated
properly. Remove the quartz sleeve and reposition for it to obtain the proper seal. Install the O-Ring down the
top of the nipple. Avoid riding the O-Ring on any threaded part of the nipple. Hand-tighten each compression
nut to form a compression seal around the quartz sleeve. Avoid overtightening the nuts, which may cause a
fracture on the end of the quartz sleeve. Under normal operation conditions, hand-tightening will provide a 100
psi seal. Do not use any devices to tighten end nuts. Repeat this procedure at the opposite end of the chamber.
After you have tightened the compression nut and all other plumbing connections, open the outlet valve. Slowly
open the inlet valve and flush out all remaining air. Then close the outlet valve and
slowly open the inlet valve fully. Check the unit for leaks. If you find a leak at the compression nut, tighten the
nut further. If the leak continues, drain the unit and inspect the quartz O-Ring and quartz sleeve for proper seal.
Once you complete checking the unit, reassemble O-Ring and tighten gland nut. Repressurize the unit and check
again.
REQUIREMENTS FOR CLEANING THE QUARTZ SLEEVE:
As water passes through the U.V., minerals, debris and other matter in the water may deposit onto the quartz
sleeve. After sufficient film has formed on the quartz sleeve, the ability of the ultraviolet
germicidal rays to pass though the quartz sleeve and into the water may be impaired. Therefore, it is necessary
to determine a cleaning schedule for the quartz sleeve. The frequency will depend on the
specific type of water conditions. If the water has been processed through deionization, reverse
osmosis, or is distilled, cleaning may be required only once per year. If untreated water is used, the cleaning
frequency will vary. A minimum of once yearly is standard recommendation for cleaning and lamp replacement.
Contact your local dealer for scheduling this service. Your specific situation will vary the frequency time accord-
ing to the water quality of the home or facility application.
QUARTZ SLEEVE CLEANING PROCEDURES:
To clean the quartz sleeve, turn off the water flow to the disinfection unit, turn power off, and disconnect the
electrical service to the lamp pins. Carefully remove the U.V. lamp. Loosen the compression nuts with O-Rings
and remove the quartz sleeve while draining the water from the chamber. The quartz sleeve may then be washed
with a mild soap and hot water solution and rinsed clean with hot water. Should this be insufficient to clean the
quartz sleeve, a mild acid may be used (i.e. vinegar). Be certain to follow all recommended safety and handling
procedures on the acid container. It is important to handle the quartz sleeve with care to prevent breakage. Make
certain that all finger prints are wiped clean before reinstalling (see installation of the quartz sleeve).
Replace the unit specific O-Rings
[P8-8-544(1), P12-8-544(1), P20-8-546(1)] every time a quartz sleeve is cleaned or replaced.