User's Manual

Table Of Contents
InnoSight Ultrasound System
M-Mode Image Controls 
 inen
94
P6992-4, EN, 16/12/6
Inverting the Color Display
Invert the color display in relation to the blood flow direction in Color mode Normally, the color red
is assigned to positive frequency shifts (flow toward the transducer), and blue is assigned to negative
frequency shifts (flow away from the transducer) Use this function to reverse this color assignment and
invert the colors on the color wedge
Touch Invert to invert the color scale
Using Directional Power
Activate Directional Power in CPA mode for use in applications where sensitivity and directional information
are both required
Touch Directional to enable this function
M-Mode Image Controls
Overview
M-Mode imaging is used simultaneously with 2-dimensional (2D mode) imaging to determine patterns
of motion for objects within the ultrasound beam M-Mode displays scan data of the anatomy in the 2D
Imaging window
a
, and the motion scan in the Time Series window
b
The M-Mode cursor line (M-line)
appears vertically in the central area of the active 2D image, indicating the position of the M-Mode beam
Typically, this mode is used for viewing motion patterns of the heart
Drag the M-line to the target position to determine the presence of motion occurred along the singular line
2D  PW
M  CW
0




Zoom 
5
 
  TGC 
2D M


 







2D


M




1

b
a



2
1

Figure 37M-Mode real-time scan
1
Touch Update to initiate the M-Mode trace which produces a scrolling display of
movement (along the vertical Y axis), plotted against time (along the horizontal  axis)
2
Touch M to enable M-Mode image controls