User's Manual
Table Of Contents
- 1 Read This First
- 2 Safety Information
- 3 Overview
- 4 Preparing the System
- Inserting a microSD Card
- Using the Stand
- Charging the System
- Connecting the Transducer
- Removing the Transducer
- Using the System On The Go
- Using the Wheel Brakes
- Placing the Transducer Holder
- Mounting the System to the System Cart
- Adjusting the System Cart Height
- Connecting an External Printer
- Tilting the System
- Unmounting the System From the System Cart
- Moving the System
- Outputting the System Display to an HDMI-Enabled TV or Monitor
- Supported External Printers
- 5 Using the System
- Turning On/Off the System
- Logging Into the System
- Launching the Main Screen
- Setting the System Time and Date
- Controlling the System
- Setting the System Language
- Identifying the Main Screen Layout
- Switching the Control Panel Pages
- Managing the System Power
- Managing Disk Space
- Network Configuration
- DICOM Configuration
- 6 Performing an Exam
- Starting a New Exam
- Adding a New Patient
- Loading a Worklist
- Selecting a Preset
- Setting the Transducer Orientation
- Selecting/Switching a Scan Mode
- Adjusting the Displayed Image
- Freezing an Image
- Adding Annotations
- Adding Measurements
- Saving and Printing the Image
- Reviewing the Image
- Exporting the Exam
- Managing the Exam List
- Ending the Exam
- 7 Using Image Controls
- 2D Mode Image Controls
- Overview
- Adjusting Gain
- Adjusting Frequency
- Adjusting Time Gain Compensation (TGC)
- Adjusting the Scan Depth
- Adjusting the Focus Depth, Focal Zone and Focal span
- Adjusting Dynamic Range
- Using Tissue Harmonic Imaging (THI)
- Adjusting Persistence
- Adjusting Sharpness and Smoothing
- Adjusting Gray Map
- Adjusting Chroma Map
- Adjusting Steer Angle
- Adjusting the Sector Width and Position
- Adjusting Power
- Using Trapezoidal Imaging
- Adjusting Density
- Using Compound Imaging
- Using ENV (Enhanced Needle Visualization)
- Color/CPA Mode Image Controls
- M-Mode Image Controls
- Spectral Doppler Mode Image Controls
- 2D Mode Image Controls
- 8 System Customization and Service
- Customizing Your System
- Configuring Security Policies
- Servicing your system
- Reinstalling Software
- Checking the Software Version
- Checking the System’s Serial Number
- Checking the Tablet’s Serial Number
- Checking the License Status
- Resetting User Settings
- Backing Up System Settings and Patient Data
- Restoring System Settings and Patient Data
- Resetting Your System
- Testing the System
- Exporting System Logs
- Reading the User Manual
- 9 Transducers
- 10 Transducer Care
- 11 System Maintenance
- 12 Appendix
InnoSight Ultrasound System
Troubleshooting
11 einenne
124
P6992-4, EN, 16/12/6
Symptom Possible cause and corrective action
The system can’t read/
write data from the USB
storage device
Two USB storage devices are connected to the system at the same time.*
» Ensure the USB storage device you wish to read/write data from is
connected to the system through the USB 3 port
The USB storage device is damaged.
» Insert the USB storage device into a computer for inspection
The USB ports are damaged.
» Insert another USB storage device into the system for inspection
No image or abnormal
display on the system
screen
The system screen is not functioning.
» Output the system display to an external monitor (See page ) and
check if images display normally on the external screen
» Contact technical support
Image Artifacts occur on
the imaging screen
Electrical interference occurs.
» Move system away from any electromagnetic sources
» Remove the power adapter and other external devices (if any) from the
system while keeping the transducer plugged for inspection
The transducer connected is damaged.
» Replace current transducer with another for inspection
» Use the diagnostic tools to test the functionality of the ultrasound engine
and the transducer (See Testing the System on page 14) This requires
assistance from technical support
Software installation
failed
Check if the USB flash drives or the system’s USB ports are damaged
The system is
overheating
The system’s fans are not functioning.
» Use the diagnostic tools to test the functionality of the fan and the
temperature (See Testing the System on page 14) This requires
assistance from technical support
Ventilation slots are blocked.
» Place the system in a well-ventilated area
» Remove any dust particles or stains found on or nearby the ventilation
slots
The Power button does
not function
The system power reaches a critically low state and is not connected to power.
» Connect the system to power
» If the power button
The system encounters
unexpected shutdowns
several times
A system disk error occurred.
» Reinstall the software (See Reinstalling Software on page 13) This
requires assistance from technical support
Touch screen is
unresponsive or
misconfigured
Touch screen is damaged.
» Inspect the panel surface carefully for cracks, cuts or any other damages
Software malfunctions.
» Connect a pointing device to the system through the USB port, and check if
the pointer is displayed correctly on the screen
» Use the diagnostic tools to test the functionality of the touch screen (See
Testing the System on page 14) This requires assistance from technical
support