User's Manual

Table Of Contents
73
 einnx
 Setting the Transducer Orientation
InnoSight Ultrasound System
P6992-4, EN, 16/12/6
Exporting and Importing Customized Presets
On the preset screen, touch Export/Import to export/import customized presets to/from your external
storage
Setting the Transducer Orientation
Upon entering all scan modes, the orientation marker ( ) displays at the default location (usually at the
top left side of the image), and suggests you the direction of holding the transducer The orientation marker
on the screen corresponds to the position of the orientation marking on the side of the transducer
Orientation marker
Figure 18Transducer orientation (Example transducer: C6-2)
You can change the left/right orientation of the image (real-time or frozen) in various imaging (single or
dual) without rotating the transducer head itself
To reverse left and right
On the imaging screen (real-time or frozen), touch
> L/R
To reverse up and down
On the imaging screen (real-time or frozen), touch
> U/D
Selecting/Switching a Scan Mode
On the real-time imaging screen, directly touch the scan mode buttons on the control panel to select/switch
the scan mode in use
To view a list of available scan modes, see Imaging on page 2
Adjusting the Displayed Image
On the real-time imaging screen, touch the corresponding scan mode (image control) button to select a
scan mode Use the image control settings to further optimize the image (See Chapter Using Image
Controls on page 9)
You can also perform the following operations to adjust the contents of the imaging window
Enlarging an Area of the Image
To further examine the anatomy by enlarging a region of the image (real-time or frozen), use the zoom
function